Operating instructions
Communication protocol Commands
KNF Flodos CP_SIMDOS02_EN_01_166851 11
9.1.1 Mode select: Run-mode / Dispense-mode
Selects the active operating mode of the pump.
MSn
MS
n (1 digit, right adjusted)
MS1 ⇒ Dispense Mode, ml and Time is active
n Function
0 Run Mode is active (Factory preset)(Auto-start
preset)
1 Dispense Mode ml and time is active
2 Dispense Mode ml/min and time is active
?MS
n (1 digit, right adjusted)
Transmit ?MS
Receive 1
n Meaning
0 Run Mode is active
1 Dispense Mode ml and time is active
2 Dispense Mode ml/min and time is active
Display on Pump User Interface:
n
Display
0 Volume: unit = ml/min Time: - - min - - sec
1 Volume: unit = ml Time: 99 min 99 sec
2 Volume: unit = ml/min Time: 99 min 99 sec
State of MS is saved
State of MS is saved
Take notice of the limits! See chapter 8.
Summary:
Set command
Command string:
Parameter name:
Parameter value:
Example:
Read command
Command string:
Answer:
Example
Power OFF behavior:
Auto-start Power OFF behavior:
Remarks: