12.4.  Status messages 
Status messages 
Description 
00 MANUAL STOP 
Circulator in „OFF“ state. 
01 MANUAL START 
Circulator in keypad control mode. 
02 REMOTE STOP 
Circulator in „r OFF“ state. 
03 REMOTE START 
Circulator in remote control mode. 
12.5.  Error messages 
Error messages 
Description 
-01 LOW LEVEL ALARM 
Low liquid level alarm. 
-02 REFRIGERATOR ALARM 
Control cable of the refrigerated circulator or MVS 
solenoid valve controller short-circuited or interrupted. 
-03 EXCESS TEMPERATURE 
WARNING 
High temperature warning. 
-04 LOW TEMPERATURE WARNING 
Low temperature warning. 
-05 WORKING SENSOR ALARM 
Working temperature sensor short-circuited or interrupted. 
-06 SENSOR DIFFERENCE ALARM 
Sensor difference alarm.  
Working temperature and safety sensors report a 
temperature difference of more than 35 K. 
-07 I
2
C-BUS ERROR 
Internal error when reading or writing the I
2
C bus. 
-08 INVALID COMMAND 
Invalid command. 
-09 COMMAND NOT ALLOWED IN 
CURRENT OPERATING MODE 
Invalid command in current operating mode. 
-10 VALUE TOO SMALL 
Entered value too small. 
-11 VALUE TOO LARGE 
Entered value too large. 
-12 TEMPERATURE MEASUREMENT 
ALARM 
Error in A/D converter. 
-13 WARNING : VALUE EXCEEDS 
TEMPERATURE LIMITS 
Value lies outside the adjusted range for the high and low 
temperature warning limits. But value is stored. 
-14 EXCESS TEMPERATURE 
PROTECTOR ALARM 
Excess temperature protector alarm 
-15 EXTERNAL SENSOR ALARM 
External control selected, but external Pt100 sensor not 
connected. 
-20 WARNING: CLEAN CONDENSOR 
OR CHECK COOLING WATER 
CIRCUIT OF REFRIGERATOR 
Cooling of the condenser is affected.  
Clean air-cooled condenser.  
Check the flow rate and cooling water temperature on 
water-cooled condenser. 
-21 WARNING: COMPRESSOR STAGE 
1 DOES NOT WORK 
Compressor stage 1 does not work. 
-22 WARNING: COMPRESSOR STAGE 
2 DOES NOT WORK 
Compressor stage 2 does not work. 