Specifications
HE  
73 
1
2
3
AK
STAND-BY 
EX-STBY: External Stand-by input 
(for external switch-off) 
Possible parameters:   
INACTIV - standby input is ignored 
ACTIV - standby input is active 
Activate the standby input: 
1.  Under menu item > EXT-STBY <, set the parameter to >ACTIV<. 
2.  Connect an external contact ‚AK‘ (e.g. for switch-off) or an alarm 
contact of the superordinated system. 
In case the connection between Pin 2 and Pin 3 is interrupted by opening 
the contact ‚AK‘, a complete shutdown of the circulating pump and heater 
is effected, and the unit enters the condition „E OFF“.  
If the contact is reclosed, the instrument returns to the standby state and 
„E OFF“ is displayed. Press 
 to start. 
AK
1
2
3
STAND-BY
Additional tips for using the STANDBY input: 
The standby function can be used in conjunction with the AUTOSTART 
feature  
1.  If the Autostart function is NOT turned ON, the standby input is used 
as described above. 
2.  If the Autostart function is enabled, the instrument will revert back to 
the original method of entering the setpoint (i.e. keypad, RS232, 
Analog input, etc.). 
Entering the setpoint with the keypad.
 e.g.    
As described above, a bipolar shutdown is accompanied by 
displaying the „E OFF“ state. The circulator starts again when the 
contact is reclosed. The temperature of the bath fluid changed during 
the „E OFF“ state. 
Entering the setpoint with the programmer.  
The display E OFF appears. The setpoint value and the time are both 
held at the current value. The temperature of the bath fluid will be held 
constant at this temperature. The programmer continues once the 
contact is reclosed. 
Notice: This is not an actual shutoff feature. 