Specifications
HE  
55 
9.5.5.  COSPEED - external 
This parameter affects the temperature pattern only in case of external 
control. 
Possible parameters: 0.0 ... 5.0 
During selftuning, the control parameters Xp, Tn and Tv of a controlled 
system are automatically determined and stored. Depending on the 
controlled system, time for tuning can be unequally longer. This 
controller layout allows protection of sensitive objects requiring 
temperature application. 
°C
t
Setpoint
5
3
0
Int. temperature
Ext. temperature
As soon as a co-speed factor is set, it is 
considered for calculating the control 
parameters. As shown in the diagram, 
tuning times become shorter the higher 
the co-speed factor is, but overshooting 
can happen in the internal system. 
9.5.6.  Control parameters – XPU-, XP-, TN-, TV- EXTERNAL 
Setting range: 0.1 ...99.9 
Setting range: 3 ...9999 
Setting range: 0 ... 999 
In most cases the control parameters preset in the factors are adequate 
for achieving an optimum temperature sequence.  
The control parameters allow adjustment to special control processes. 
Setting range: 0.1 ... 99.9 
Proportional range >Xpu< 
The proportional range Xpu of the cascaded controller is only needed for 
external control. 