Specifications
Operator responsibility – Safety recommendations 
18 
2.5.  Cooling water connection 
Cooling water pressure (IN / OUT )   max.  6 bar 
Difference pressure (IN - OUT )    3.5 to 6 bar 
Cooling water temperature    <20 °C 
Recommended quality of cooling water: 
pH – value  
7,5 to 9,0 
Sulfate [SO4 2- ]  
< 100 ppm 
Hydrocarbonate [HCO3- ] / Sulphate [SO4 2-]  
> 1 ppm 
Hardness [Ca2+, Mg2+] / [HCO3-]  
> 0,5 dH 
Alkalinity  
60 ppm < [HCO3-] < 300 ppm 
Conductivity  
< 500 μs / cm 
Chloride (CL-)  
< 50 ppm 
Phosphate (PO43-)  
< 2 ppm 
Ammonia (NH3)  
< 0,5 ppm 
Free Chlorine  
< 0,5 ppm 
Ferri Ions (Fe3+ )  
< 0,5 ppm 
Mangano Ions (Mn2+)  
< 0,05 ppm 
Carbon dioxide (CO2)  
< 10 ppm 
Hydrosulfide (H2S)  
< 50 ppm 
Content of oxygen  
< 0,1 ppm 
Algae growth  
impermissible 
Suspended solids  
impermissible  
Notice: 
Danger of corrosion of heat exchanger due to unsuitable quality of cooling 
water. 
•  Due to its high content of lime hart water is not suitable for cooling and causes 
calcination of the heat exchanger. 
•  Ferrous water or water containing ferrous particles will cause formation of rust 
even in heat exchangers made of stainless steel. 
• 
Chlorous water will cause pitting corrosion in heat exchangers made of stainless 
steel. 
•  Due to its corrosive characteristics distilled and deionized water is unsuitable 
and will cause corrosion of the bath. . 
•  Due to its corrosive characteristics sea water is not suitable. 
•  Due to its microbiological (bacteria) components which settle in the heat 
exchanger untreated and unpurified river water and water from cooling towers is 
unsuitable. 
•  Avoid particulate matter in cooling water. 
•
  Avoid putrid water. 
Notice:   Cooling water circuit 
Risk of oil leaking from the cooling circuit (compressor) of the recirculating cooler into 
the cooling water in case of a fault in the circuit! 
Observe the laws and regulations of the water distribution company valid in the 
location where the unit is operated. 