User manual
152 MicroMAX System User Manual Version 6.C
SPEX 500M (NTE with or without shutter)
4
1
5
2
3
Qty P/N Description
1. 1 2517-0214 Adapter-Female, Spex 500m
2. 8 2826-0170 Screw, 1/4-20 × 0.51, Low Socket Head Cap, Black
3. 1 2518-0223 Adapter-Male, Spex 500m
4. 3 2826-0134 Screw, 10-32 × 1/4, Flat Head Slot, Stainless Steel
5. 2 2826-0055 Screw, 8-32 × 14, Set Allen Hex, Nylon
Assembly Instructions
1. Insert part 1 into the spectrometer wall (dashed line in illustration) and fasten with
the socket head screws provided.
2. Attach part 3 to the camera with the three (3) 1/4" long button head screws provided.
3. Gently insert part 3 into part 1 and fasten with the setscrews.
Note: Adapter parts are machined to provide a tight fit. It is necessary to rotate the camera
back and forth when inserting into the spectrometer adapter. Do not force the two parts of
the adapter together, as they can be permanently damaged by excessive force.