Operating instructions
Code English output Description
15 Digital Input Malfunction at the digital input
20 T ext Ethernet No signal for actual value via the Ethernet module
8.3
Low Level alarm
If the liquid level falls below the minimum level, an alarm
sounds.
Low Level is shown on the screen. The device components
such as the pump are switched off by the electronics.
Personnel:
Operating personnel
1. Refill lacking heat transfer liquid.
2. Unlock the display with the OK button.
The device restarts.
8.4
Overtemperature alarm
This alarm can only occur for devices with heater.
The message 3 Overtemperature is shown on the screen if the
temperature monitor trips.
The electronics switch off the components of the device.
The device outputs a two-tone signal.
Fig. 72: Alarm Low Level
Rectifying fault
Fig. 73: Overtemperature alarm
Faults