Operating instructions
Setting Description
Tolerance Tolerance defines the level of accuracy with
which the end temperature should be
reached before the next segment will be pro-
cessed.
S1, S2, S3 Switching contacts of the contact module (if
available) can be programmed here. Contact
modules are available as accessories.
The graph shows an example of reprogramming a setpoint tem-
perature profile.
Fig. 59: Program example
The cooling time in the graph varies depending on device type,
consumer etc. In the sample segment
no. 2, 50 °C should be
reached within 20 minutes.
The original values of the following table, “previously“, are shown
with solid lines, the edited profile of the following table, “after-
wards“, with a dashed line.
previously (
____
)
No. Tend hh mm Tol Pum
p
S1 S2 S3
Start 30.00 -- -- 0.1 --- off off off
2 50.00 0 20 0.0 --- off off off
3 70.00 0 40 0.0 --- off off off
4 70.00 0 10 0.1 --- off off off
5 60.00 0 30 0.0 --- off off off
6 30.00 0 0 0.0 --- off off off
A new segment with the number 3 was entered in the edited table.
The time for the segment with number 4 was also changed. The
tolerance for the segment with number 5 was adjusted.
Operation
Variocool 63