Operating instructions

Personnel:
Operating personnel
1. Change to the main menu.
2.
Select the menu item Setup Basic setup Language.
3. Select any of the available languages.
4. Confirm with the OK button.
6.12
Entering the offset of the internal actual temperature (calibration)
Deviations from calibrated reference thermometers such as
LAUDA DigiCal can be corrected internally using the "Offset" func-
tion.
Personnel:
Operating personnel
1. Change to the main menu.
2.
Select the menu item Setup Calibration Calibration.
3. Adjust the value accordingly. The value displayed on the ref-
erence thermometer must be entered.
4. Confirm with the OK button.
6.13
Restoring factory calibration (internal temperature sensor)
Any offset specified for the internal temperature measurement can
be restored to the factory setting.
Personnel:
Operating personnel
1. Change to the main menu.
Fig. 47: Selecting menu language
Fig. 48: Specifying offset
Operation
Variocool 58