Operating instructions
Personnel:
Operating personnel
1. Change to the main menu.
2.
Select the menu item Setup Basic setup Autostart.
3. Select one of the following options:
With off , the device is in the standby operating mode
after a mains power failure.
Normal operation is continued directly after a power
failure using on .
4. Confirm with the OK button.
6.11.5
Limiting current consumption
If your line fuse is below 16 A, the current consumption can be
reduced in steps from 16 A to 8 A. In the case of the optional
heater, the output of the heater is reduced accordingly. Take
account here whether other consumers are still connected to the
same fused circuit or whether your device is the only consumer.
Personnel:
Operating personnel
1. Change to the main menu.
2.
Select the menu item Setup Basic setup
Curr. Consumpt.
3. Adjust the current consumption accordingly.
4. Confirm with the OK button.
Fig. 43: Specifying Autostart
Fig. 44: Specifying current consump-
tion
Operation