User guide
Page A-2 P/N. 11016-001 Rev. January 7, 2005
User Guide
AutoPulse™ Resuscitation System Model 100
A.5 Platform Environmental
Table A-3 Environmental Specifications
Category Specifications
Operating temperature +32º to +104ºF (0º to +40ºC).
Storage temperature -4º to +149ºF (-20º to +65ºC).
Relative humidity 5% to 95%, non-condensing.
Atmospheric pressure 0 to 15,000 feet above sea level (760 to 428 mmHg).
Water resistance Splash resistant as defined by IP22 per International Electrotechnical
Commission (IEC) 60529.
Safety classification Meets IEC 60601 – internally powered equipment, Type BF-Defibrillation
Protected, movable, short-time operation, Class III.
Electromagnetic
susceptibility
IEC61000-4-3, 4, 5, and 6 – level 2 (80 MHz to 2 GHz, 3V/m).
Electrostatic discharge Meets IEC 61000-4-2 – 6 KV Contact, 8 KV Air.
Electromagnetic emissions Meets CISPR 11/EN55011, Group 1, Class A.
Patient contacting materials Meets ISO 10993-1 Biological evaluation of medical devices.
Shock Meets IEC 60068-2-27 Basic Environmental Testing – Shock (50g, 11ms pulse,
half sine wave).
Vibration
• Meets IEC 60068-2-64 Basic Environmental Testing – Random Vibration
Broad Band (f1:20-f2:2000, ASD: 0.05).
• Meets IEC 60068-2-6 Environmental Testing – Vibration (sinusoidal), (10
to 150 Hz, 10m/s
2
).
Drop Meets IEC 60068-2-32 Basic Environmental Testing – Free Fall – Procedure 1.
Corrosion resistance External components are non-corrosive.
Operating classification Short-time per IEC 60601-1 (30 minutes).