User guide

Page 4-2 P/N. 11016-001 Rev. January 7, 2005
User Guide
AutoPulse™ Resuscitation System Model 100
Figure 4-2 Battery Charger Control Panel
4. To manually run a test-cycle on a Battery, simply press the Battery Charger’s Start Test button at any
time that the Battery Charger is in Charge mode.
Note: Do not remove a Battery from the Battery Charger until its charging completes or the
Battery’s runtime will be reduced.
Note: Do not remove a Battery during a test-cycle or the Battery’s runtime will be unknown.
Removing a Battery during a test-cycle may cause the Battery Charger to automatically
enter a test-cycle mode the next time a Battery is inserted into it.
5. Remove the Battery from the Battery Charger by gripping the Battery while pressing the finger latch
to disengage the locking bar. Pull the Battery straight out until it fully clears the charging bay.
Note: Recently-charged Batteries can be warm to the touch. This is a consequence of normal
operation.
4.2 Cleaning the AutoPulse Platform
1. Remove and dispose of the LifeBand Chest Compression Assembly (CCA).
2. Wipe all the surfaces of the AutoPulse Platform free of foreign matter and spills with a disinfectant
or bactericidal wipe. Check the vents to ensure that they are free and clear of any obstructive matter.
Caution: Do not submerge the AutoPulse Platform in liquid.
3. Ensure that the AutoPulse Platform is dry before storing.
4.3 Storing the AutoPulse Platform
1. The AutoPulse Platform should be powered down before it is stored.
Note: The AutoPulse Platform will automatically power down after 10 minutes of inactivity.
2. The AutoPulse Platform should be returned to its storage case, or a dust cover used, before being
stored.
3. The AutoPulse Platform should be stored in a cool, dry place.
4.4 Maintenance
There are no user-serviceable parts in the AutoPulse Platform. In case of repair or service, contact the
manufacturer.