User guide

P/N. 11016-001 Rev. January 7, 2005 Page 3-9
User Guide
AutoPulse™ Resuscitation System Model 100
4. Press the Start/Continue button a second time to start compression cycles. The AutoPulse Platform
begins chest compression cycles (see Figure 3-11).
Note: Do not lean on the patient after pressing the Start/Continue button.
Figure 3-11 Chest Compression Display Panel Screen
5. Depending on the Mode setting in Administrative Menu (refer to section Section 2.3,
Administrative Menu: user pre-set options,” on page 2-9), the AutoPulse Platform will perform
15:2 or Continuous compressions. In 15:2 mode the Platform performs 15 compressions and then
pauses for three seconds to permit the user to ventilate the patient before automatically resuming
compressions (see Figure 3-12). In Continuous mode the Platform performs continuous
compressions. If on-the-fly mode switching has been enabled (in the Mode setting within the
Administrative Menu) then the AutoPulse Platform will perform in the mode (either 15:2 or
Continuous) that was used last until powered down; on power up 15:2 will be the initial selection.
In 15:2 mode three audio cue tones will sound prior to the ventilation pause: one during each of the
13th, 14th, and 15th compressions. In Continuous mode, in every 7 compressions, one audio cue
tone will sound—on the 6
th
compression. The tones can be temporarily disabled (and re-enabled)
by pressing the Tone Silence button (refer to section Section 1.4.2.6, “Tone Silence Button,” on page
Warning:
If you must move or realign the patient, you must press the Stop/Cancel button before
adjustment.
Do not place your hands or any other objects on or under the LifeBand chest bands while the
AutoPulse System is analyzing the patient or during active operation.