User guide

P/N. 11016-001 Rev. January 7, 2005 Page 3-7
User Guide
AutoPulse™ Resuscitation System Model 100
Figure 3-8 Fastening the LifeBand Chest Bands
3.2 Starting Chest Compressions
1. Make sure that the yellow upper edge of the LifeBand chest bands are aligned with the patient's
armpits, and is directly over the yellow line on the AutoPulse Platform. Also make sure that there
are no obstructions, such as clothing, straps or equipment, with the bands.
Warning:
Failure to properly position the LifeBand Chest Compression Assembly (CCA) at the
patient’s armpit line may cause injury to the patient.
Failure to properly position a patient, both vertically and laterally with respect to the
AutoPulse Platform, may cause injury to the patient.
Do not strap across, or otherwise constrain, the LifeBand chest bands. Constraining the
movement of the bands can damage or break the LifeBand CCA.
Operating the AutoPulse System on a patient for extended periods of time may result in
minor skin irritation to the patient.