AutoPulse™ Resuscitation System Model 100
Figure 2-12 Pause Alert Tones menu
The “Set Pause Alert Tone” menu item allows you to set the frequency or disable the tone sequence used
to alert that the system has been deliberately stopped (paused) while actively doing compressions. The
options are to set the system to beep once every “60 seconds” (one minute), “30 seconds” or “15
seconds.” Highlight the desired setting using the Move Up and Down arrow buttons, press the Select
Choice Button to select it (a check will appear beside the selected item) and then press the gray Menu/
Mode switch button under the word “back” to return to the main Administrative Menu.
Note: If the pause alert tone is disabled here it cannot be re-enabled, during opperation, by using the
Tone Silence button (refer to section Section 1.4.2.6, “Tone Silence Button,” on page 1-7).