User guide

Page 2-10 P/N. 11016-001 Rev. January 7, 2005
User Guide
AutoPulse™ Resuscitation System Model 100
To exit from the Administrative Menu, press the “START” (green) button under the word “RESTART.
The AutoPulse System will restart and place you into the idle state, ready for patient alignment or for
system power-down.
Note: The current setting is displayed in the parenthesis after the main menu item.
Figure 2-10 Compression Mode menu
The “Set Mode” menu item allows you to restrict the AutoPulse Platform operation to a single
compression mode or to allow on-the-fly mode switching. Selecting “15:2 or Continuous” will allow on-
the-fly mode switching between 15:2 and continuous compressions while the system is actively doing
compressions. Selecting “15:2 Only” will restrict the system operation to the 15:2 mode. Selecting
“Continuous Only” will restrict the system operation to continuous compressions. Highlight the desired
setting using the Move Up and Down arrow buttons, press the Select Choice Button to select it (a check
will appear beside the selected item) and then press the gray Menu/Mode switch button under the word
“back” to return to the main Administrative Menu.