AutoPulse™ Resuscitation System Model 100
Figure 2-7 Removing the LifeBand Chest Band from the AutoPulse Platform
4. Remove the LifeBand chest bands from the AutoPulse Platform by grasping the LifeBand chest
band with the thumb and index finger of both hands, on either side of the LifeBand band clip.
Push in the guide plate, using both middle fingers. Keeping the guide plate pushed in,
pull up the band to remove the clip from the shaft (see Figure 2-7).
5. Discard the LifeBand CCA as it is a single-use component after patient use. Treat the
LifeBand CCA as contaminated medical waste and dispose of it accordingly. There are no
user-serviceable parts.
6. Inspect the new, replacement LifeBand CCA for cuts or tears.
7. Replace the LifeBand CCA following the procedures in Section 2.1.1, “Installing the LifeBand
CCA”.