User guide
P/N. 11016-001 Rev. January 7, 2005 Page 2-1
User Guide
AutoPulse™ Resuscitation System Model 100
2 Preparing the AutoPulse System for Use
The AutoPulse System is delivered fully assembled, except for the LifeBand Chest Compression
Assembly (CCA) and the AutoPulse Power System Battery.
2.1 LifeBand Chest Compression Assembly (CCA)
Installation and Removal
2.1.1 Installing the LifeBand CCA
1. Place the AutoPulse Platform with the patient surface facing down on a smooth, flat surface.
Note: The drive shaft should be oriented so that the slot faces directly upward.
2. Insert the head end of the LifeBand band clip into the drive shaft slot. The correct direction is
towards the LifeBand plate alignment arrow seen on the Platform (see Figure 2-1).
Figure 2-1 Sliding the LifeBand Band Clip into the Drive Shaft Slot
3. Once the head end of the LifeBand band clip is positioned into the slot, press the tail end of
the band clip into the slot of the guide plate until the band clip is fully seated in the drive shaft. You
should feel it lock into place.
4. Make sure that the band clip is seated properly and fully into the slot on the drive shaft (see Figure
2-2).
Note: If the band clip is properly seated, you should be able to easily turn the drive shaft in each
direction by hand.
1
LifeBand Cover Plate
LifeBand plate alignment arrow
LifeBand Band Clip
Tail End
Head End
Slotted Drive Shaft
Guide Plate
1
2
2