User guide

P/N. 11016-001 Rev. January 7, 2005 Page 1-11
User Guide
AutoPulse™ Resuscitation System Model 100
1.4.4 Performance Characteristics
The basic operating characteristics of the AutoPulse Platform are shown in Table 1-3. All performance
characteristics are factory preset. No adjustments are necessary.
Table 1-3 Operating Characteristics
Operating Performance Specification
Compression rate 80 (± 5 compressions per minute)
Compression Modes (user selectable)
15:2 (15 compressions with two 1.5 second ventilation pauses)
Continuous compressions
Duty cycle 50 (± 5%)
Compression depth 20% of chest depth, +0.25/-0.5 inch