User guide
P/N. 11016-001 Rev. January 7, 2005 Page 1-9
User Guide
AutoPulse™ Resuscitation System Model 100
1.4.2.8 Power (Green LED)
The green Power LED lights whenever the AutoPulse Platform is powered on and able to respond to user
input.
1.4.2.9 Alert (Red LED)
The red Alert LED lights whenever a user advisory, fault or system error condition exists for the
AutoPulse Platform. For a list of advisory, fault and error conditions, refer to Chapter 5,
“Troubleshooting Procedures”.
1.4.3 Battery Charge Status
The User Control Panel displays the battery charge status. The Battery charge status icon only appears
when the AutoPulse Platform is powered up.
Figure 1-5 Low Battery Warning
When at least five minutes of active operation remain on a Battery, the User Control Panel will give a
“Low Battery” indication (see Figure 1-5). The “Low Battery” indication will remain on until the Battery
is replaced or depleted. The Low Battery warning display will be accompanied with an audio warning of
four rapid beeps which will be followed by two beeps every 30 seconds until the battery is replaced or
depleted. It is recommended that a fully-charged Battery be exchanged for the Battery with the low
charge (if available).
Indicates the level of charge of the Battery. A graphic battery icon indicating four
proportional levels of battery charge is displayed.
15:2
LOW BATTERY