User guide
P/N. 11016-001 Rev. January 7, 2005 Page 1-7
User Guide
AutoPulse™ Resuscitation System Model 100
1.4.2.5 Select Choice Button
1.4.2.6 Tone Silence Button
Pressing the Select Choice button selects the currently highlighted menu or list
item.
c
Selecting the Tone Silence button will disable or enable audio (tone) feed back
generated by the AutoPulse System. When the audio feedback is on, pressing the
Tone Silence button will turn it off. When the audio feedback is off, pressing the
Tone Silence button will turn it back on. If ventilation and/or pause tones have
been disabled in the Administration Menu the Tone Silence button will not re-
enable them (refer to section Section 2.3, “Administrative Menu: user pre-set
options,” on page 2-9).
Once audio feedback has been disabled with the Tone Silence button, it will
automatically be re-enabled when one of the following occurs:
1. You press the Tone Silence button again.
2. You press the Start/Continue button to begin compressions.
3. You press the Stop button during compressions.
4. A low battery condition is reached (refer to section Section 1.4.3, “Low
Battery Warning,” on page 1-9).
5. You switch between compression modes (refer to section Section 3.2,
“Mode Change Confirm Display Panel Screen,” on page 3-11).
The icon displayed on the User Control Panel display when
tones are enabled (on).
The icon displayed ion the User Control Panel display when
tones are disabled (off).