10. PACKAGING, STORAGE AND HANDLING
Pulsante® SPG Microstimulator
The Microstimulator is provided sterile. The Microstimulator is sterilized using ethylene oxide.
Do not excessively bend the Microstimulator Lead or Bone Fixation Plate. Excessive bending
may damage the integrity of the Microstimulator and may affect device performance.
Do not attempt to reuse or re-sterilize the Microstimulator. The Microstimulator is intended for
single use only.
Do not use the Microstimulator if the current date is beyond the labeled “Use by” date.
Do not use the Microstimulator if the packaging or its contents are damaged or have been
previously opened.
Do not store the Microstimulator packaging where it may be exposed to water or other liquids,
as moisture may damage the seal integrity of the package materials.
Store the Microstimulator between 10°C (14°F) and +55°C (+131°F); temperatures outside
of this range may damage components.
Pulsante® Surgical Tools
The Surgical Tools are provided sterile. The Surgical Tools are sterilized using ethylene oxide.
Do not attempt to reuse or re-sterilize the Surgical Tools after surgical use. The Surgical Tools
are intended for single use only.
Do not use the Surgical Tools if the packaging or its contents are damaged or have been
previously opened.
Do not store the Surgical Tools’ packaging where it may be exposed to water or moisture.
Pulsante® Remote Controller
Do not sterilize the Remote Controller.
Do not store the Remote Controller where it may be exposed to water or other liquids, as
moisture may damage the integrity of the package materials or damage the Remote Controller.
Do not use the Remote Controller if the packaging or the Remote Controller look damaged.
The Remote Controller is a sensitive electronic device that may be damaged by rough
handling, including dropping on the ground or being crushed.
The Remote Controller can be safely stored and operated between 5°C and 40°C. The
Remote Controller can be transported between 20°C and +70°C. Temperatures outside of this
range may reduce battery life.
Do not connect the USB Port to any other electrical device other than the Charger provided
with the Remote Controller or the Programming USB cable provided with the Programmer
Software.
The Remote Controller may feel warm when providing therapy. If the patient experiences
discomfort, recommend moving the Remote Controller away from the face slightly and avoid
contact with skin.
The Remote Controller should not come into contact with skin that is cut or bleeding.
Clinician Programmer
Do not sterilize the Clinician Programmer.
Do not store the Clinician Programmer where it may be exposed to water or other liquids, as
moisture may damage the integrity of the package materials or damage the Clinician
Programmer.
Do not use the Clinician Programmer if the packaging or the Clinician Programmer look
damaged.