Instruction manual

SM425 INSTRUCTION MANUAL 2.0 SPECIFICATIONS
2.7.2 Dust Guard Protector (Part No.: K9471UC)
Recommended when process gas flows directly into the measurement cell, flammable particulate flows
into the measurement cell, or water droplets may leak into the measurement cell during downtime.
Material: SUS316
Weight: Approx. 0.3 kg
Figure 2-11 Dust guard protector dimensions
2.7.3 Auxiliary Ejector for High Temperature (Part No. E7046EC or E7046EN)
This ejector is used where pressure of measured gas for high temperature detector is negative. This
ejector consists of an ejector assembly, a pressure gauge and a needle valve.
Standard Specifications
Ejector Assembly
Ejector Inlet Air Pressure: 29 to 68 kPa G
Air Consumption: Approx. 30 to 40 l/min
Suction gas flow rate: 3 to 7 l/min
Connection to high-temperature probe adapter: R 1/2
Connection: Rc1/4 or 1/4 FNPT, SUS304 (JIS)
Tube Connection: (Ø6/Ø4 mm or 1/4 inch copper tube or stainless tube)
Pressure Gauge Assembly
Type: JIS B7505, A1.5U3/8 x75
Material in Contact with Gas: SUS316 (JIS)
Case Material: Aluminum alloy (Paint color; black)
Scale: 0 to 100 kPa G
Bushing (G3/8 x Rc1/4 or 1/4NPT, SUS304 (JIS))
Needle Valve
Connection: Rc1/4 or 1/4FNPT
Material: SUS316 (JIS)
30 42500001 REV A