2.0 SPECIFICATIONS SM425 INSTRUCTION MANUAL
Table 2-1.
External Dimensions - refer to Figure 2-1.
2.3.2 ZO21P-H Adapter for High-Temperature Probe
The probe adapter is used to lower the sample gas temperature to below 700°C (below 300°C at probe
adapter surface) before it is fed to the detector.
Insertion length:
1 m or 1.5 m
Material in Contact with Gas:
SUS 316 (JIS), Zirconia, SiC or SUS 310S, SUS 304(JIS) (flange)
Probe Material:
SiC, SUS 310S (JIS)
Installation:
Flange mounting (FF type or RF type)
Probe Mounting Angle:
Vertically downward within ± 5°
Where the probe material is SUS 310S, horizontal mounting is available.
Construction:
Non explosion-proof. Rainproof construction
Weight:
Insertion length of 1.0 m: approx. 6.5 kg (JIS) / approx. 8.5 kg (ANSI)
Insertion length of 1.5 m: approx. 7.5 kg (JIS) / approx. 9.5 kg (ANSI)
Sample gas temperature:
0 to 1400°C (when using SiC probe) 0 to 800°C (when using SUS 310S probe adapter)
Sample gas pressure:
-0.5 to + 5 kPa. When using in the range of 0 to 25 vol% O
2
or more, the sample gas pressure
should be in the range of - 0.5 to + 0.5 kPa. (Where the sample gas pressure for the high-
temperature probe is negative, an auxiliary ejector is necessary.)
Table 2-3 High temperature probe adapter codes
42500001 REV A 13