Instruction manual

SM425 INSTRUCTION MANUAL 2.0 SPECIFICATIONS
12 42500001 REV A
2.3 High-Temperature Separate-type Detector and Related
Equipment
2.3.1 SM425P (0.15m) High-Temperature Separate-type Detector
Standard Specifications
Construction:
Water-resistant, non-explosion proof
Probe length:
0.15 m
Terminal box:
Aluminum alloy
Probe material:
Probe material in contact with gas: SUS 316 (JIS) (Probe), SUS 304 (JIS) (Flange), Zirconia
(Sensor), Hastelloy B, (Inconel 600, 601)
Weight:
Approx. 3 kg
Installation:
Flange mounting (The use of high-temperature detector probe adapter ZO21P-H is necessary.)
Flange standard:
JIS 5 K 32 FF equivalent (thickness varies)
Mounting angle:
Any angle between the horizontal and vertical (high-temperature probe is fitted with an adapter)
Reference gas and calibration gas piping connection: Rc 1/4 or 1/4 NPT female
Cable inlet:
G 1/2, Pg 13.5, M20 3 15, 1/2 NPT
Ambient temperature:
-20 to 150°C
Sample gas temperature:
0 to 700°C (temperature at the measuring point of the sampling gas), 0 to 750°C, or 0 to 1400°C
when the probe adapter for high temperature is used.
Temperature of the probe adapter shall not exceed 300°C to protect the gasket and avoid the
bolts seizing together.
Sample gas pressure:
- 0.5 to + 5 kPa: when used at the range of more than 0 to 25 vol% O
2
, - 0.5 to + 0.5 kPa. (An
auxiliary ejector is required for negative pressure application.)
Model and Code
Refer to “Model and Code” in