Instruction manual

SM425 INSTRUCTION MANUAL 12.0 TROUBLESHOOTING
If an Error-2 occurs immediately after power is supplied, the polarity of the thermo-couple input
connection (TC+, TC-) in the converter may be reversed. Check detector connections.
Possible causes of an Error-2 condition are shown below.
1. Faulty heater in the detector (heater wire breakage)
2. Faulty thermocouple in the detector
3. Faulty cold junction sensor located at the detector terminal block.
4. Failure of the converter electronics
5. Heater temperature control overcurrent limiting triggered.
6. TC+ and TC- thermocouple terminals wired to detector with reverse (wrong) polarity.
Overcurrent protection is triggered if there are problems in the heater wiring. When the protective
circuit is triggered, an internal fuse blows. The heater is then disconnected, resulting in an Error 2
(temperature failure) condition.
<Troubleshooting and Corrective Actions>
1. Turn off power to the converter.
2. Remove the cable from terminals 7 and 8 of the detector and measure the resistance value
between these terminals. The heater unit will be normal if the resistance is lower than about 90.
If the resistance value is higher, failure of the heater unit is suspected. In this case, replace the
heater unit (refer to Section 1.59.3, “Replacement of the Heater Unit”). In addition, check that the
wiring resistance between the converter and detector is 10 or less.
3. Ensure that TC+ terminal (terminal 3 in detector) is connected to converter TC+ terminal, and TC-
terminal (terminal 4) is connected to converter TC- terminal.
4. Remove the wiring from terminals 3 and 4 of the detector and measure the resistance between
these terminals. The thermocouple is considered normal if the resistance value is 5 or less. If
the value is higher than 5, damaged thermocouple wiring may be the cause. In this case,
replace the heater unit (refer to Section 1.59.3, “Replacement of the Heater Unit”). Also, check
the wiring resistance between the converter and detector is 10 or less.
5. Even if items (2) to (4) are normal, the heater overcurrent protection fuse may have blown. Check
for wiring problems such as the following:
Heater terminals are shorted together.
Heater terminal(s) are shorted to ground.
Heater terminals are shorted to power supply.
Do not attempt to replace an open internal fuse. Contact your Teledyne Monitor Labs service
representative if this occurs.
• Measure the thermocouple resistance after the difference between the detector tip temperature and the
ambient temperature decreases to 50°C or less. Accurate resistance measurements cannot be made at
high temperatures.
190 42500001 REV A