8.0 DETAILED DATA SETTING SM425 INSTRUCTION MANUAL
Table 8-5 Output contact states
8.4.2 Programming the Output Contacts
Only output contacts 1, 2, and 3 can be programmed. Each of these contacts can be activated by multiple
alarms and “other conditions”. The alarm or condition is selected by turning it “ON”. In addition, the state
of the active contact can be programmed to be open or closed.
To program the output contacts, follow these steps:
1. Press the Setup key in the basic panel display screen to display the Execution/Setup display
screen.
2. Select Setup in the Execution/Setup display screen. The “Commissioning” (Setup) display screen
appears.
3. Select the Contact setup in the “Commissioning” (Setup) display screen. The Contact setup
display shown in Figure 8-7 appears.
4. Select the desired output contact. In this example contact output 1 has been selected (see Figure
8-8).
5. Table 8-6 lists the alarms, errors, and conditions that can be used to activate an output contact.
6. Select “Others” in the Output contact 1 display screen. The “Contact1” display shown in Figure
8-9 appears. Select Calibration in the Contact1 display screen.
7. The “ON/OFF” selection display appears. Select “ON”.
8. Press the Return key to go back to the previous display.
9. Select “During power-off the contact is open and in condition (when the contact is active) it is
Open”. The “ON/OFF” display appears. If “OFF” is selected, the contact will be open in normal
conditions and closed when the contact output is activated.
• Contact output 4 can not be programmed.
42500001 REV A 127