8.0 DETAILED DATA SETTING SM425 INSTRUCTION MANUAL
8.2 Configuring Output Hold
The “output hold” function retains an analog output signal at a predetermined value during certain
equipment states. Outputs 1 and 2 can be set independently. Table 8-2 shows the output hold selections
available for each of the states.
Table 8-2 Output Hold Selections
Equipment states During warm-
up
Under
maintenance
Under calibration
During blow back
On Error
Occurrence
(*1)
Output hold
selections
available
4 mA
○
20 mA
○
No hold feature
○ ○ ○
Retains output
from just before
occurrence
○ ○ ○
Preset value
(2.4 to 21.6 mA)
○ ○ ○ ○
○: The output hold functions are available
*1: The output hold functions for on error occurrence are unavailable when option code “/C2” or “C3” (NAMER NE 43 compliant) is
specified.
8.2.1 Definition of Equipment States
1. During warm up: “During warm up” is the time required after applying power until the sensor
temperature stabilizes at 750°C, and the equipment is in the measurement mode.
2. Under maintenance: “Under maintenance” is the time starting when the Setup key on the Basic
Panel display screen is pressed until returning back to the Basic Panel display screen. The sub-
screens under the Execution/Setup display screen are included. See Figure 7-4.
3. Under calibration (see Section 0, Calibration): For manual calibration “Under calibration” is the
time starting from entering the Manual calibration display screen (shown in Figure 8-1) until the
Calibration End key is pressed and the preset output stabilization time elapses.
Figure 8-1 Manual calibration
42500001 REV A 121