7.0 STARTUP SM425 INSTRUCTION MANUAL
Note all settings will be reinitialized to default settings if the converter type selection is changed.
To confirm the converter type, follow these steps:
1. Press the setup key.
2. Use the key to select Setup and press the [Enter] key.
3. Press the [Enter] key on the password entry display screen if a password has not been set for
commissioning. Otherwise enter the password for the Commissioning display screen. See
Section 1.50.5, “Setting Passwords” for information on setting a new password.
4. Select “Basic setup” and press the [Enter] key on the Commissioning display screen as shown in
Figure 7-8.
5. The Basic setup display screen shown in Figure 7-9 then appears. Confirm “This instrument is for
Oxygen” is displayed.
6. If the converter type is to be changed, press the [Enter] key. The display screen shown in Figure
7-10 appears.
7. Use the key to select the type of equipment. Then press the [Enter] key to complete the
converter selection.