Section 6 MC-III™ EXP Flow Analyzer
9. Mount the display/keypad assembly to the enclosure with the two #4-40 × 7/8” screws.
10. Recalibrate the MC-III EXP if necessary and replace the enclosure cover.
Firmware Update
Before installing new rmware or moving the instrument to a new site, it is recommended that the user
clear the EEPROM, which will erase all conguration settings and log data. To clear the EEPROM, choose
Options/Program Options from the menu bar, select Advanced from the dropdown options list, and click on
the “Clear EEPROM” button (Figure 6.7).
Technical assistance is typically required for upgrading rmware. Contact Cameron’s Measurement Systems
Division to arrange an upgrade.
!
WARNING: Clicking the “Clear EEPROM” button will delete all conguration settings and log
records in the MC-III EXP memory.
Figure 6.7—EEPROM clearing and automatic data log option