User manual
96
Section 6 MC-III™ EXP Flow Analyzer
Figure 6.5—Disassembly of circuit board/keypad assembly
8. Disconnect the keypad ribbon cable from the J3 connector on the LCD side of the circuit assembly as fol-
lows:
a. Grasp the black clip between a thumb and forenger (see white arrows in Figure 6.6 below).
b. Squeeze both sides of the clip and gently pull to release the clip from the plastic connector that holds
it in place. DO NOT PULL on the ribbon cable. When the black plastic clip is properly disengaged,
the cable will release freely.
Figure 6.6—To release the ribbon cable from the connector, press in on the side tabs of the connector (white
arrows) and gently pull forward (black arrow).