Section 2 MC-III™ EXP Flow Analyzer
Pulse Output
The pulse output is a solid-state relay. Each pulse represents a user-dened volume. Because the circuit is
isolated, it can be used in conjunction with any other feature on the MC-III EXP. A two-conductor cable from
the MC-III EXP to the remote location is required (Figure 2.13). The maximum current rating of the pulse
output circuit is 60 mA at 30 VDC.
A Zener diode (Part No. 1.5KE56CA) must be installed for CE approval.
For information on conguring the pulse output with the keypad, see Conguring the Pulse Output, page 43.
For information on conguring the pulse output with the interface software, see Conguring Pulse Output,
page 73.
TB1
PULSE
INPUT
RESET
INPUT
TFM
A&S
GND
T
B
3
EXT POWER
TB2
RS485
SLAVE
4-20
OUT
PULSE
OUT
6
-
3
0
V
D
C
J1
J2
RESET
SWITCH
BATTERY
Leave this
end of shield
disconnected.
POWER SUPPLY
5 to 30 VDC
*
Resistor may be included in pulse readout
device. Size the resistor to limit the current
to 60 mA.
*
PULSE READOUT
DEVICE
Figure 2.13—Pulse output wiring