MC-III™ EXP Flow Analyzer Appendix D
Real Time
Register
(Decimal)
Register
(Hex)
Description
Data
Type
Access
1200 04B0
Year
(Real Year = register value plus 2000)
U16 R/W
1201 04B1 Month [1-12] U16 R/W
1202 04B2 Day [1-31] U16 R/W
1203 04B3 Hour [0-23] U16 R/W
1204 04B4 Minute [0-59] U16 R/W
1205 04B5 Second [0-59] U16 R/W
This block of registers is used to set the instrument’s internal clock. To set the time, it is recommended that all
registers be written in one message.
The time and date can also be read in the holding register groups as oating-point data.
Input Conguration
Register
(Decimal)
Register
(Hex)
Description
Data
Type
Access Default
2000 07D0
Volume Units
Units
0 = No Units
1 = Gallons
2 = Barrels
3 = Cubic Meters
4 = Cubic Feet
5 = Liters
Units × 1000
16 = No Units
17 = Gallons
18 = Barrels
19 = Cubic Meters
20 = Cubic Feet
21 = Liters
U16 R/W 2
Standard Units
32 = No Units
33 = Gallons
34 = Barrels
35 = Cubic Meters
36 = Cubic Feet
37 = Liters
Std. Units × 1000
48 = No Units
49 = Gallons
50 = Barrels
51 = Cubic Meters
52 = Cubic Feet
53 = Liters
2001 07D1
Volume Decimal Point Location
1 = Tenths
2 = Hundredths
3 = Thousandths
4 = Ten Thousandths
U16 R/W 1