Service manual

Model 5.J04A
Maintenance
Table 9D-5-1. In-Cabinet Perfomrance Checks (Continued)
6.
Totalizing (OPEN/CLOSE
A)
Measurements.
Obtain
the
following test equipment:
HP
Model 65IB Test Oscillator.
HP
Model U048B 50-0hm Feed-thru
Termination.
a.
Connect
651B
50-0hm output,
set to
10 Hz at 100 mV rms, to
5304A Chan-
nel A
INPUT connector.
b. Set
Channel
A
ATTEN switch
to X100,
AC-DC
to AC;
SLOPE
to and
LEVEL
toPSET.
c. Set
"Function" switch
to
OPEN/CLOSE
A.
d.
Instrument gate
is
controlled
by
oper-
ator when OPEN/CLOSE
A
pushbutton
is
pressed.
e.
Press OPEN/CLOSE
A
switch;
ft MX) A C
lamp will
tum on;
adjust Channel
A
ATTEN
to Xl;
5300A display should
accumulate.
f. To
stop 5300A from accumulating, press
OPEN/CLOSE
A. To
continue accumu-
lation from
the
number displayed press
OPEN/CLOSE
A,
otherwise press
RESET
to
clear display.
90-5-19