Service manual

Modei 5304A
Maintenance
OPEN/CLOSE MODE
5304A Controls:
Channel
A, SEP, XI, DC, +,
PSET
OPEN/CLOSE
A.
Press RESET, then
OPEN/CLOSE; display should
be ac-
cumulating
at a 10 Hz
rate.
INPUT:
651B set to 10 Hz, 100 mV
rms/50<2.
Oscilloscope:
HP 180A, HP 1801A
Vert
Amp, HP 1821A T.B.
through
a 10:1
probe.
Settings:
AC
coupled (except
as
noted).
NORM sweep mode.
+
slope (except
as
noted)
AC
(except
as
noted)
Channel
A
5300A Sample Rate
ccw out of OFF
(except
as
noted).
5304A: OPEN/CLOSE Mode
TPl H.
Goes
Low
when OPEN/CLOSE switch
is
pressed. Reamains Ix>w
if
this switch
is
held depressed.
TP2 L.
Goes High when RESET
is
pressed.
Re-
mains High
if
this switch
is
held depressed.
TP3 L
TP4 L.
Goes High when OPEN/CLOSE
is
pressed
to
stop accumulation
or
when RESET
is
pressed during accumulation.
TP5 H.
Goes
Ix)w
when OPEN/CLOSE
is
pressed
to
stop accumulation
or
when RESET
is
pressed during accumulation.
9D-5-12