Service manual
Model 5304A
Maintenance
PERIOD AVERAGE
A
MODE
5304A Controls:
Channel
A, SEP, XI, AC, +,
PSET,
PER AVG A.
INPUT:
651B set to 1 kHz, 100 mV
rms/5012.
Oscilloscope:
HP 180A, 1801A
Vert
Amp, 1821A T.B.
through
a 10:1
probe.
Settings:
AC
coupled (except
as
noted).
NORM sweep mode.
+
slope (except
as
noted).
Channel
A.
5300A
SAMPLE RATE
ccw out of OFF
(except
as
noted). 5300 Display should
be
approximately
1.00000
MS, C.
HP
10525A Logic Probe
Test points using
H =
High (lamp
on)
L = Low
(lamp
off)
TPl H
(dim). Flash rate decreases when 5300 SAMPLE
RATE
is
turned
cw.
Goes
Low
when RESET
is
pressed.
TP2 H
(dim). Flash rate decreases when 5300 SAMPLE
RATE
is
turned
cw.
Goes Ix>w when RESET
is
pressed.
TP3 H
(dim). Flash rate decreases when 5300 SAMPLE
RATE
is
turned
cw.
Goes High (bright) when
RESET
is
pressed.
TP4 H
(dim). Flash rate decreases when 5300 SAMPLE
RATE
is
turned
cw.
Goes High (bright) when
RESET
is
pressed.
TPS L.
Goes High when RESET
is
pressed.
TP6 H
TP7 H
9D-5-6