Service manual
Model 5304A
Maintenance
FREQ
A,
AUTO MODE.
5304A controls:
Channel
A, SEP, XI, AC, +,
PSET, AUTO
INPUT:
HP 651B set to 10 MHz, 100 mV
rms/50S2.
Oscilloscope:
HP 180A, 1801A
Vert.
Amp. 1821A T.B.
through
a 10:1
probe.
Settings:
AC
coupled (except
as
noted).
NORM sweep mode.
+
SLOPE (except
as
noted).
ACF
(except
as
noted).
Channel
A.
5300A SAMPLE RATE
ccw out of OFF
(except
as
noted). 5300 Display should
be
approximately
the
same
as the HP 651B
setting.
Test points using
HP
10525A Logic Probe
H =
High (lamp
on)
L = Low
(lamp
off)
TPl H
(dim). Flash rate decreases
as
5300
SAMPLE RATE control
is
truned
cw.
TP2 H
(dim). Flash rate decreases
as
5300
SAMPLE RATE control
is
turned
cw.
I^evel goes
Low
when 5300A RESET
is
pressed.
TP3 H
(dim). Flash rate decreases
as
5300A
SAMPLE RATE control
is
turned
cw.
Level goes
Low
when 5300A RESET
is
pressed.
TP4 H
(dim). Hash rate decreases
as
5300A
SAMPLE RATP] control
is
turned
cw.
Level goes High when 5300A RESET
is
pressed.
TP5 H
(dim). Flash rate decreases
as
5300A
SAMPLE RATE control
is
turned
cw.
Level goes High when 5300A RESET
is
pressed.
TP6 L.
Goes High when
53(H)
R&SET
is
pressed.
TP7 L.
Goes High when 5300A RESET
is
pressed.
TP8
9D-5-4