Service manual
Model 5304A
Operation
SECTION
IX D
5304A TIMER/COUNTER
SUBSECTION
III
OPERATION
9D-3-1. OPERATING INFORMATION
9D-3-2.
The
5300A/5304A Timer/Counter measures
frequencies
up to 10 MHz,
period average
to 1 MHz,
time interval from
500
nsec
to
10,000 seconds;
or
totalizes input signals.
Figure 9D-3-1. Signal Conditioning Using
Attenuator, Level
and
Slope Controls
TRIGGER
VOLTAGE
NOTE
?
RECOVERY
VOLTAGE
+
SLOPE
0
LEVEL
PSET
HYSTERESIS
(NOTE
I)
-
SLOPE
I.
HYSTERESIS (NOISE REJECTION)
VO.
TAGE
IS
CONSTANT
FOR
EACH ATTENUATOR SETTING
AND
DOES
NOT
VARY WITH
TRIGGER LEVEL
2
THIS ALLOWS ZERO VOLTS TRIGGERING WITH GOOD NOISE
REJECTION
3
ATTENUATOR:
xi,
xio.xioo CHANGES TRIGGER LEVEL
AND
HYSTERESIS VOLTAGES
BY
FACTORS
OF
I.IO.IOO
9D-3-3. There
are
four basic signal conditioning
controls used
in
Channel
A or
Channel
B (the
effect
of
these controls
is
shown
in
Figure 9D-3-1).
These
are:
a.
ATTEN switches
for
Channel
A and B.
b.
SLOPE switches
for
Channel
A and B.
c.
LEVEL controls
for
Channel
A and B.
d.
AC-DC switches
for
Channel
A and B.
9D-3-4. ATTEN.
XI, X10,
XI00 switches (both
Channels. Large amplitude signals
can be
attenuated
by
using
the
ATTEN switches. Setting
the
attenuator
switches increases
the
trigger levels
and the
hysteresis voltages
by
factors
of 1, 10, 100. If
input
signal levels
are
unknown,
the
initial measurements
should
be
made with
the
attenuator switches
set
to
XI00.
The
attenuator settings
are
then reduced
until
a
stable, useable measurement
is
obtained.
9D-3-5. SLOPE Switches (both Channels).
The
slope switches allow
the
selection
of
triggering
on
the
positive
(+) or
negative
(-)
slope
of the
input
signals
(see
Figure 9D-3-1).
9D-3-6. LEVEL Controls (both Channels).
The
level controls allow adjustment
of the
triggering
point
on the
input signal waveform. With
the
LEVEL
controls
set to
PSET
or "0",
triggering
is
centered
about zero volts.
The
voltage range over which
triggering
may be set is -1
volt
to +1
volt times
the
ATTEN switch settings
(±1 V to +100 V).
9D-3-7. AC-DC Switches (both Channels).
The
AC-DC switches select
ac or dc
coupling.
In AC
position
and
PSET, triggering
is
centered about
ac-zero,
and dc
components
on the
signal have
no
effect.
In DC
position
and
PSET, triggering
is
centered about
0
volts
dc on the
signal.
9D-3-1