Service manual

Model 5304A
General Information
Table 91)1-1. 5304
A
Timer/Counter Specifications
INPUT CHANNELS
A ANI) B
Range:
DC
coupled;
0 to 10 MHz.
AC
coupled;
100 Hz to 10 MHz.
Sensitivity (min):
25 mV rms
sine wave
to I MHz. 50 mV rms
sine wave
to 10
MHz. 150 mV p-p
pulse
at
minimum
pulse width,
40
nsec. Sensitivity
can be
decreased
by 10 or 100
times using
ATTENUATOR switch.
Impedance:
1 Mil
shunted
by
less than
30 pK
Overload Protection:
250 V rms on Xl()
and
XI00 attenuator settings.
On XI
attenuator setting
120 V rms up to 1 kHz,
decreasing
to 10 V rms at 10 MHz.
Trigger Level: PRESET position centers
triggering about
0
volts
±25 mV, or con-
tinuously variable over
the
range
of 1 V to
+
1 V
times attenuator settings.
Slope: Independent selection
of
triggering
on
positive
or
negative slope.
Channel Inputs: Common
or
separate lines.
Gate Output: Rear panel
BNC. TTL low
level while gate
is
open.
May be
used
to
intensity modulate
an HP
oscilloscope.
TIME INTERVAL
Range:
500
nsec
to 10 sec.
Input: Channels
A and B; can be
common
or
separate
Overload: Maximum recommended levels
for
pulse signals
is 2 V p-p.
Clipping occurs
at 3 V p-p.
Resolution:
100
nsec
to 10 ms in
decade
steps.
Accuracy:
+1
count
+
time base accuracy
+
trigger error.*
Display:
fjs, ms, or s
(seconds) with positioned
decimal point.
Time Interval Holdoff: Front panel
con-
centric knob which inserts variable delay
of
approximately
100
/jsec
to 100
msec
be-
tween START (Channel
A) and
enabling
STOP (Channel
B); may be
disabled. Elec-
trical inputs during delay time
are
ignored.
Delay
may be
digitally measured
in
CHECK
position. Delay Output: rear panel
BNC.
TTL low
level during delay time.
May be
used
to
intensity modulate
an HP
oscilloscope.
PERIOD AVERAGE
Range:
10 Hz to 1 MHz.
Input: Channel
A.
Period Averaged:
1 to 10'
automatically
selected
for
maximum resolution.
Frequency Counted:
10 MHz.
Accuracy:
+
1
count
+
time base accuracy
+
trigger error.**
Display:
/JS, ms
with positioned decimal point.
FREQUENCY
Range:
0 to 10 MHz.
Input: Channel
A.
Gate Times: Manually selected
0.1, 1, or 10
seconds. AUTO position selects gate time
to
fill display
for
maximum resolution with-
in a
1-second measurement time.
Accuracy:
+1
count
+
time base accuracy.
Display:
Hz, kHz, and MHz
with positioned
decimal point.
OPEN/CLOSE (Totalizing)
Range:
10 MHz
maximum.
Input: Channel
A
Function: Input signal totalized while gate
open. Opening
and
closing
of
gate initiated
by
front panel pushbutton switch.
GENERAL
Check: Inserts internal
10 MHz
reference
frequency into Channels
A and B.
Time
Interval Holdoff
may be
digitally measured
by
switching
to
CHECK
and
TIME
IN
TERVAL positions.
Operating Temperature:
0° to
50°C.
Power Requirements: Including 5300A
mainframe, nominally
10
watts. Minimum
of 3
hours operation (typically
5 hrs) at 20 C
to
30°C operating
and
charging temperatures.
Weight:
Net, 1 lb (0.9 kg).
Shipping,
314
lbs (1.5 kg).
"For any
waveshape, trigger error
is
less than
±
0.005
Signal Slope
(V/ ps) /js
**Trigger error
is
less than
±0.3% of one
period,
i
periods averaged
for
signal with
40 dB or
better signal-to-noise ratio.