OPERATING AND SERVICE MANUAL TIMER/COUNTER 5304A w.
CERTIFICATION The Hewlett-Packard Company certifies that this instrument was thoroughly tested and inspected and found to meet its published specifications when it was shipped from the factory. The HewlettPackard Company further certifies that its calibration measurements are traceable to the U.S. National Bureau of Standards to the extent allowed by the Bureau's calibration facility. WARRANTY A N D A S S I S T A N C E This Hewlett-Packard product is warranted against defects in materials and workmanship.
III' M< >1 >EL5;MMA TIMER colINTER Reeord of Revisions (Manual Change I'ages) REVISION NUMBER DATE ISSUED DATE INSERTED INSERTED BY
HI' MODEL :">;{( M A TIM E R CO IJNTK K Record of Revisions (Manual Change Pages) REVISION NUMBER I DATE ISSUED DATE INSERTED INSERTED BY
it SECTION IX D TIMER/COUNTER 5304A SERIAL PREFIX: 1212A00467 This section applies directly to HP Model f>H()lA, Timer Counters having serial prefix number l2P2A004ti7, and must he inserted into the 5300A Measuring System M uiual. NEWER INSTRUMENTS This Section with enclosed "Manual Changes Pages" applies directly to HP Model 5.'U)4A Timer C o u n t e r s having prefix numbers above 1212A00467.
Model 5304 A Table of Contents TABLE OF CONTENTS Section IX I) i'age 5304A Timer/Counter Subsection ii I GENERAL INFORMATION 91)-1-1. Introduction 91)-1-2. Description 91)-1-4. Purpose and Use of Section IX I) 91)1-6. Instrument Identification 91)1-8. Manual Changes and Options 91)1-1 9D-1-1 9D-1-1 91)-1-1 91)1-1 91)1-1 II INSTALLATION 9D-2-1. Unpacking and Inspection 9D-2-3. Storage and Shipment 9D-2-5. Installation and Removal of Plug-On 9D-2-8.
Model 5304A List of Figures List of Tables LIST OF F I G U R E S Figure Page 91)1-1. 5304A Timer/Counter 9D-1-0 9D-3-1 9D-3-2. 9D-3-3. 9D-3-4. 9D-3-5. 9D-3-6. 9D-3-7. 9D-3-8. Signal Conditioning Using Attenuator, Level and Slope Controls Front-Panel Controls and Connectors Rear-Panel Connectors Making Self-Check Measurement Making Frequency A Measurements Making Period Average Measurements Making Time Interval Measurements Totalizing 9D-3-1 9D-3-2 9D-3-4 9D-3-5 9D-3-H 9D-3-7 9D-3-8 9D-3-10 9D-5-1.
Model 5304A General Information Figure 91)-1-1.
Model 5.'{(MA ( ieneral Informatinn S E C T I O N IX D 5304A TIMER/COUNTER SUBSECTION I GENERAL INFORMATION 9 D - 1 - 1 . INTRODUCTION 9D-1-6. INSTRUMENT IDENTIFICATION 9 D - 1 - 2 . Description 91)1 -7. Hewlett-Packard uses a two-section, ninedigit serial number (0000A00000) mounted on the rear panel to identify the instrument. The first four digits are the serial prefix and the last five digits refer to the specific instrument.
Model 5304A General Information Table 91)1-1. 5304 A Timer/Counter Specifications INPUT CHANNELS A ANI) B R a n g e : DC coupled; 0 to 10 MHz. AC coupled; 100 Hz to 10 MHz. PERIOD AVERAGE R a n g e : 10 Hz to 1 MHz. I n p u t : Channel A. S e n s i t i v i t y (min): 25 mV rms sine wave to I MHz. 50 mV rms sine wave to 10 MHz. 150 mV p-p pulse at minimum pulse width, 40 nsec. Sensitivity can be decreased by 10 or 100 times using ATTENUATOR switch. + Accuracy: 1 count + time base accuracy + trigger error.
Model 5304 A Installation S E C T I O N IX D 5304A TIMER/COUNTER S U B S E C T I O N II INSTALLATION 9D-2-1. UNPACKING A N D INSPECTION 9D-2-2. If the shipping carton is damaged ask that the carrier's agent be present when the instrument is unpacked. Inspect the instruments for damage such as scratches, dents, broken knobs, etc.
OPERATO IN ( j
Model 5304A Operation S E C T I O N IX D 5304A TIMER/COUNTER S U B S E C T I O N III OPERATION 9 D - 3 - 1 . OPERATING INFORMATION 9D-3-2. The 5300A/5304A Timer/Counter measures frequencies up to 10 MHz, period average to 1 MHz, time interval from 500 nsec to 10,000 seconds; or totalizes input signals. Figure 9D-3-1. Signal Conditioning Using Attenuator, Level and Slope Controls 9D-3-3.
Model 5304A Operation Figure 9D-3-2. Front-Panel Controls and Connectors T.I. A TO B 10 m« / PEP, v AVO/A FREQ A \ XlOO PUT LEVE AYTEN 0 0 ^ .1 ms DO OPEN.XCLOSE A K g ] 5304A TII/ER/COUNTER 1. COM-SEP-CHK. Three position switch enables selection of input signal from separate or common sources or enables the selection of a Self-Check position. a. COM: Connects INPUT A and INPUT B i n parallel. b. SEP: Channel A and Channel B inputs are from separate sources. c.
Model 5304A Operation Figure 9D-3-2. Front-Panel Controls and Connectors (Continued) 4. AC-DC. Two-position switch allows selection of ac-coupling or direct-coupling to input amplifiers. 5. SLOPE +, -. Selects the triggering on positive or negative slope of the input signals. 6. LEVEL. Adjusts the input trigger levels. Preset to trigger at 0 volts. 7. FUNCTION. Twelve-position switch enables 4 modes of operations to be selected. The modes are: c.
Model 5304A Operation Figure 9D-3-3 Rear-Panel Connectors GATE OUT DELAY OUT 1. GATE OUT: A TTL-LOW level signal while the 5300A main gate is open. 2. DELAY OUT: A TTL-LOW level signal during the delay time selected by frontpanel DELAY control (in T.I. A to B mode only).
Model 5304A Operation Figure 9D-3-4. Making Self-Check Measurements l(REAR) / I A TO B FREQ A PER XIO 10 m s xioo I INPUT PSE ATTEN SLOPE I ms LEVEL 10 p s XIO I MS XiOC » mi .1 m s 0 XI DC PSET** l us 5 3 0 4 A TIMER/COUNTER 1 Connect ac power to 5300A ac receptacle. 2. Turn ac power "on" with 5300A SAMPLE KATE control. Adjust SAMPLE KATE for desired display time. 3. Set 5304A COM SEP-CHK switch to CHK. 4. Set 5304A "Function" switch to FREQ A, AUTO. Display should he 10.
Model 5304A Operation Figure 9D-3-5. Making Frequency A Measurements I (REAR) / T . I . A TO B PER 10 ms v AVG ' t ms PSET^ INPUT ATTEN SLOPE LEVEL Q + —^ \ • FREQ A AUTO It .1 ms I s 10 |iS 10 8 I ps MS XIOO — OPEN IMER/ 1. Connect ac power to 5300A ac receptacle. 2. Turn ac power "on" with 5300A SAMPLE KATE control and adjust sample rate for desired display time. 3. Set COM-SEP-CHK to SEP. 4. Connect input signal to INPUT A jack. 5.
Model 5304A Operation Figure 9D-3-6. Making Period Average Measurements I (REAR) FREQ A NPUT xioo m DC • 53/4A TIMER/COUN/ 1. Connect tic power to 5300A ac receptacle. 2. Turn ac power "on" with 5300A SAMPLE KATE control. Adjust sample rate for desired display time. 3. Set COM-SEP-CHK to SEP. 4. Connect input signal to INPUT A jack. 5. Set A I T EN to XI00, AC-DC to coupling desired and SLOPE to polarity desired. Set LEVEL to PSET. ().
Model 5304 \ Operation Figure 9D-3-7. Making Time Interval Measurements l(REAR) AUTO 1. Connect ac power to 5300A ac receptacle. 2. Turn ac power "on" with 5300A SAMPLE RATE control. Adjust sample rate for desired display time. 3. Set "Function" switch to T.I. A to B, with desired resolution. Ensure DELAY control is full ccw. 4. Set ATTEN switches for both channels to XI00; AC-DC switches and SLOPE controls switches to settings desired; set LEVEL controls to PSET. 5.
Model 5304A Operation Figure 9D-3-7. Making Time Interval Measurements (Continued) b. If the input signals are pulse trains, it is possible to select a pulse out of the train by increasing the DP]LAY from full ccw; the measured time interval will increase in steps as the delay falls after successive pulses in the train. HP 182!A: Set TRIGGER to EXT. SLOPE to -. Coupling to ACF. SWEEP MODE to NORM. Time/CM for one cycle of measured Time Interval. c.
Model 5304A Operation Figure 9D-3-8. Totalizing l(REAR) SAMPLE RATE GSD 5300A MEASURING SYSTEM HEWLETT • PACKARD' "1 0 PSET' INPUT ATTEN SLOPE * T.I. A TO 8 10 m s FREQ A PER AVG A , \ ' V 4* AUTO LEVEL XI00 [Mi OPEN OPENXCLOSE 5 3 0 4 * TIMER 1. Connect ac power to 5300A ac receptacle. 2. Turn ac power "on" with 5300A SAMPLE RATE control. Adjust sample rate for desired display time. 3. Set "Function" switch to OPEN ' CLOSE A. 4.
THY
Model 5304A Theory of Operation S E C T I O N IX D 5304A TIMER/COUNTER S U B S E C T I O N IV THEORY OF OPERATION 9 D - 4 - 1 . INTRODUCTION 9D-4-2. This subsection describes the theory of operation for the 5304A Timer/Counter. Basic operation of gates, certain amplifiers and integrated circuits is found in Section IV of the 5300A portion of the manual. 9D-4-3. To simplify measurement making refer to operating information starting with Paragraph 9D-3-1 for optimum adjustment of various controls.
Model 5304A Theory of Operation the 5300A m a i n f r a m e at A1 Pi(16) is routed through U12C as the F2 signal to control the time base, which clocks the opening and closing of the main gate. 9D-4-15. The positive-going edge of a negative pulse from U17E is the MAX TIME signal. The MAX TIME signal a t A1 Pi(17) triggers a display cycle in the 5300A mainframe. This negative-going pulse can be generated from one of two sources.
Model 5304A Theory of Operation signal from the 5300A A1J 1(18) is gated through Q18, U17F, U9B, and U16B to be counted (except in the .1 fjsec position, where the 10 MHz clock is gated through U12A and U16B). The TIME BASE OUTPUT signal is a 10 MHz clock signal divided down by the time base to allow counting in increments of 1 fjsec, 10 fjsec etc. 9D-4-26. The main gate closes when Channel B "clocks" U15A and a display cycle is initiated by the positive-going edge of the M G F F signal at A1P1(12).
MAN I TENANCE
Model 5304A Maintenance S E C T I O N IX D 5304A TIMER/COUNTER SUBSECTION V MAINTENANCE 9 D - 5 - 1 . INTRODUCTION 9D-5-2. This section contains maintenance and service information for Model 5300A/5304A Timer/ Counter. Included are performance check procedures and tests to localize, isolate, and locate defective components. 9 D - 5 - 3 . RECOMMENDED TEST E Q U I P M E N T 9D-5-4.
Model 5304A Maintenance Figure 9D-5-1.
Model 5304 A Maintenance performance checks indicate a malfunction or if instrument operation is suspect, perform the SelfCheck procedures in Table 9D-5-1. For further tests, separate the 5304A from the casting and reconnect to 5300A using Paragraph 9D-5-6 steps a to e as a guide. Operating procedures in Subsection 9D-3 can be used to help understand operation. To test the 5304A, obtain the test equipment listed in Table 5-1 of the 5300A portion of the manual. 9D-5-15.
Model 5304A Maintenance FREQ A, AUTO MODE. Test points using HP 10525A Logic Probe H = High (lamp on) 5304A controls: L = Low (lamp off) Channel A, SEP, XI, AC, +, PSET, AUTO INPUT: TPl H (dim). Flash rate decreases as SAMPLE RATE control is truned cw. TP2 H (dim). Flash rate decreases as 5300 SAMPLE RATE control is turned cw. I^evel goes Low when 5300A RESET is pressed. TP3 H (dim). Flash rate decreases as 5300A SAMPLE RATE control is turned cw. Level goes Low when 5300A RESET is pressed.
O O .01 V/cm .1 \i sec/cm o .2 V/cm .1 (isec/cm dc coupled 1 V/cm lfisec/cm .01 V/cm O .l|J.sec/cm .2 V/cm . 1 (isec/cm dc coupled © . 1 V/cm lOmsec/cm dc coupled - slope .01 V/cm O . lfisec/cm .1 V/cm .1 fj-sec/cm © .2 V/cm 10 msec cm dc coupled - slope .2 V/cm .2fj.sec/cm dc coupled - slope .2 V/cm . 1 fJ-sec/cm dc coupled ® © .1 V/cm .1 fisec/cm .2 V/cm .
A2 ATTENUATOR BOARD ASSEMBLY QD Ai TIME INTERVAL BOARD ASSEMBLY OSKX-SOOQI: 60CQ?I JD 016 013 Qtva *j j B1 • 0 0 B 0 • 017 PA) AlPI 1 OlO 03 1" 09 2 r 02 <0 0 06 06 " CM) _5i5
&TQR BOARD ASSfMBlY ">M04 «oc fuse) 6 —f>" i»"53TW •>9 <5»T*t 410 enw > O OiAKtCL B SLOPE a E C T O H [j^»> * j B ;• U ' 4 A j Lt*> S' D>G'T A0CWES5 "*"!)7 Jr- o :> • [§3 0 0 0 0 0047 a 0 SfLECl I—[->»' OiGlT 51G 'I O MMi CLOCK . OUU#«L SLOPE SELECTION I OPEN 'CLOSE * 1 C A N S T v.
Model 5304A Maintenance HP 10525A Logic Probe PERIOD AVERAGE A MODE Test points using 5304A Controls: H = High (lamp on) Channel A, SEP, XI, AC, +, PSET, PER AVG A. INPUT: L = Low (lamp off) TPl 651B set to 1 kHz, 100 mV rms/5012. H (dim). Flash rate decreases when 5300 SAMPLE RATE is turned cw. Goes Low when RESET is pressed. Oscilloscope: HP 180A, 1801A Vert Amp, 1821A T.B. through a 10:1 probe. Settings: AC coupled (except as noted). NORM sweep mode. + slope (except as noted). Channel A.
O .01 V/cm 1 msec/cm ® .01 V/cm 1 msec /cm 0 .05 V/cm 50 jisec/cm dc coupled O .2 V/cm .5 msec/cm dc coupled O .2 V/cm .5 msec/cm dc coupled © .2 V/cm . 1 fjLsec cm dc coupled .2 V/cm .1 |.isec/cm dc coupled O .2 V/cm 10 msec/cm dc coupled - slope .2 V/cm .2fj.sec/cm dc coupled - slope 0 .2 V/cm .2 fxseccm dc coupled - slope .2 V/cm 10 msec cm dc coupled slope A 5 msec/cm dc coupled - slope 10 msec cm dc coupled O n .2 V/cm .
Ai TIME INTERVAL BOARD ASSEMBLY iosw< toooM A2 ATTENUATOR BOARD ASSEMBLY u m o * 6000?) -4 DraQ tm •D O O J ® ^ *-0 o- 015 QJ 013 016 014 f Ll vT.
A> A S S E M B L Y (09J04 6000?! TIME I N T E R V A L BOARD ASSEMBLE •-coo ;_OIT .015 016 01 oie [ LtVU" * "1 | LEVU • 1 Q 0 Q Q 0 0 @ , S ls£3 !o2 •Qi7 f * t «.:. si * j P / 0 AlPI 010 ©6 V 6 r]) +sv— 09 OH • w.
Model 5304A Maintenance No. 2 HP 222A: TIME INTERVAL A TO B Mode, .1 (isec position. Connect Equipment connections: shown in Figure 9D-5-4A. equipment REP RATE to MAN/EXT +. as PULSE WIDTH to .05-.5. Vernier adjusted for oscilloscope display of .1 fisec. Control Settings: 5300/5304A. PULSE POLARITY to +. PULSE AMPLITUDE for a oscilloscope. ATTEN (both channels) to XI. AC-DC (both channels) to DC. SLOPE (both channels) to +. LEVEL.
Model 5304A Maintenance Part of Figure 9D-5-4. Time Interval Troubleshooting Diagram .1 V/cm .1 [isec/cm .2 V/cm .1 jisec/em dc coupled - Slope .05 V/cm . 1 fisec/cm dc coupled - Slope o .2 V/cm . 1 fisec/cm dc coupled .2 V/cm .1 fisec/cm dc coupled - Slope .05 V/cm . 1 |j.8ec/cm - Slope .05 V/cm . 1 fisec/cm dc coupled - Slope .1 V/cm .1 fj.sec/cm .05 V/cm . 1 fisec/cm .2 V/cm . 1 p. sec/cm dc coupled .2 V/cm .
Model 5304A Maintenance Part of Figure 9D-5-4. Time Interval Troubleshooting Diagram © .2 V/cm . 1 (isec/cm dc coupled Q A © .2 V/cm .2fjisec/cin dc coupled - Slope .2 V/cm .1 fisec. <-m dc coupled Q H .2 V / c n 5(i8ec/cm dc coupled - Slope (DELAY IN DETENT) 9D-5-10 o 0 .2 V/cm .1 jj.8ec/cm dc coupled .2 V/cm .1 msec/cm dc coupled - Slope (DELAY just out of DETENT. 5300 Display and waveform should be the same) ® A Q B .2 V/cm 5p.sec/cm dc coupled - Slope (DELAY IN DETENT) .2 V/cm .
© H .5 V/cm 50 |j.sec/cm dc coupled (I)BLAY just out of DETENT. .5300A Display time should be approximately onehalf the oscilloscope waveform time) © © A .5 V/cm 50 fisec/cm dc coupled (DELAY IN DETENT) © .1 V/cm 50 fj.sec/cm dc coupled .5 V/cm 50 jisec/cm dc coupled (DELAY just out <>f DETENT. 5300A Display should be approximately 200 > to 240|isec) © A .2 V/cm .1 [isec/cm dc coupled • Slope © © A .2 V/cm 20 fisec/cm dc coupled - Slope (DELAY just out of DETENT.
A! TIME INTERVAL BOARD ASSEMBLY tos* A2 ATTENUATOR BOARD ASSEMBLY
Al TIME I N T E R V A L HOAHD A S S t M b L Y (0S304-60000 | ^ cuts'*, B SlDPt St- ECTlON 1 ] 1 ^*1 * I ] B 0 0 • ) ' tOC Ol/TPOT ^ <* OPT*I • iO K55E O 1 i ) J6 J'011 400Ht5s 1—rr_l^s f 16 ri >0 UHl CLOCK OVU#«l A SLOW S£LECTiON <0 , [ 0P1N/CLOSE * 1 S f S W E2I9 - j JK <#H 8 INHIBIT CHANNEL A S W U C H 1 7. ? l TIME BASE INPUT 131 u z o 6 .
Modei 5304A Maintenance OPEN/CLOSE MODE 5300A Sample as noted). Rate ccw out of OFF (except 5304A Controls: Channel A, SEP, XI, DC, +, PSET OPEN/CLOSE A. Press RESET, then OPEN/CLOSE; display should be accumulating at a 10 Hz rate. INPUT: 5304A: OPEN/CLOSE Mode TPl 651B set to 10 Hz, 100 mV rms/50<2. Oscilloscope: HP 180A, HP 1801A Vert Amp, HP 1821A T.B. through a 10:1 probe. Settings: AC coupled (except as noted). NORM sweep mode.
o .01 V/cm 20 msec/cm ® .02 V/cm 20 msec/cm O .05 V/cm 50 msec/cm dc coupled ® .2 V/cm 50 msec dc coupled © .2 V/cm 50 msec/cm dc coupled O .2 V/cm 50 msec/cm dc coupled O .
AI TIME INTERVAL BOARD ASSEMBLY it A? ATTENUATOR BOARD ASSEMBLE « m 0 4 6Qoo?i 015 El ED > C2; m • • 014 016 I LtVtl * | I ItVCL B 1 • 0 • • 0 Q E«D,5 ® " » 017 P/0 AlPl -r-6 + I7V 010 OOljiF €5 <> 1 J) 7V 05 — 06 08 012
Al TIME I N T E R V A L BOARD ASSEMBLY IOSSO*-6000.: ,DELA* Or ^9 (5PTSi ^ 10 5.T55T } a TIME BASE OUT DIGIT SELECT OANNEL e SLOPE SELECTION I LEVEL * | 9 £V I LEVEL B | • 0 0 • 0 0 @ , g S D go l u ' W )i2 OAW«I M5R * SLOPE SELECTION SI •HI ( / • , , — CHANNEL * SWITCH- —' \ ,, " 1 6 12 UIZO P/C A l P I I CHANNlL EXPONENT C O U N T E H Ei(51 TiMJ BASE INPUT •V M DIGIT SELECT \ M DIGIT AOOBESS ® rr*t IIUtMJ -;^3 ' •, * B SWITCH EXPONENT STOWAGE — .
Model 5304A Maintenance CHECK MODE 5304A Controls: C O M / S E P / C H K to CHK; "Function" to AUTO 5300A count. Display should be 10.0000 MHz C ±1 Oscilloscope: HP 180A, HP 1801A Vert Amp, HP 1821A T.B. through a 10:1 probe Settings: AC coupled (except as noted). NORM sweep mode. + Slope (except as noted). AC (except as noted). Channel A.
o • .2 V/cm • * M-8ec/cm decoupled O .2 V/cm . 1 fj.sec/cm dc coupled .2 V/cm 1 fisec/cm decoupled .2 V/cm .
A2 ATTENUATOR BOARD ASSEMBLY COSJO*-coooz) Al TIME INTERVAL BOARD ASSEMBLE EJ y0'5 fl-n 016 014 013^ • 00 00B Isp.
Al TIME INTERVAL BOARD ASSEMBLY
Model 5304A Maintenance Table 9D-5-1. In-Cabinet Performance Checks 1. CHK mode and Time Interval Holdoff. 2. FREQ A Mode. RANGE: dc coupled: 0 to 10 MHz. a. Set 5300A OSC switch to INT. b. Mate 5300A to 5304A and ensure ie casting latches at rear of 5300A are latched. Sensitivity as in Table 9D-1-1. c. Connect ac power to 5300A ac receptacle. Obtain following test Equipment: d. Turn ac power "on" with 5300A SAMPLE RATE slightly cw out of OFF.
Model 5304A Maintenance Table 9D-5-1. In-Cabinet Performance Checks (Continued) Figure A Figure H HP S300A HP 2??A HP iBuA HP 5300A HP 6510 HP i BOA HP 5304A CHAN A V t R T IN PULSE OUTPUT H P 5304A TO C H A N A CHAN A VERT IN NPUT 10MHz soft FEEDTHRU FEEDTHRU oNC TEE b. BNC TEE Set 222A for "+" PULSE POLARITY; 40 nsec pulse width; 150 mV p-p as monitored on oscilloscope. c. Monitor 222A pulse output with oscilloscope and set REP RATE for 10 Hz. d.
Model 5304A Maintenance Table 9D-5-1. In-Cabinet Performance Checks (Continued) , n . Z AXIS . INPUT i Figure C * 1 L pJ I DELAY • • OUT l l HPM30A HP ??2A I80IA 30IA AT PULSE OUTPUT 50A FEEOTHRUS 1 INPUT 182 IA 182IA , f BED T INPUT A INPUT D SOft FEEOTHRUS fl II BNC TEE HP 222A EXT TRIG INPUT b. DELAYED PULSE PULSE OUTPUT A BNC TEE While monitoring generator No. 1 output with oscilloscope, set No. 1 generator controls as follows: f. 5304A should display 00000.
Model 5.J04A Maintenance Table 9D-5-1. In-Cabinet Perfomrance Checks (Continued) 6. Totalizing (OPEN/CLOSE A) Measurements. Obtain the following test equipment: c. Set "Function" switch to OPEN/CLOSE A. d. Instrument gate is controlled by operator when OPEN/CLOSE A pushbutton is pressed. e. Press OPEN/CLOSE A switch; ft MX)A C lamp will tum on; adjust Channel A ATTEN to Xl; 5300A display should accumulate. f. To stop 5300A from accumulating, press OPEN/CLOSE A.
Model 5304A Maintenance PERFORMANCE CHECK TEST CARD Hewlett-Packard Model 5300A/5304A TIMER/COUNTER Serial No - Test Performed by Date Description 1. CHECK Mode and Time Interval Holdoff. Check • Display is 10 MHz ±1 Count. Hold off as in Table 9D-5-1, item 1, step i. 2. F R E Q A M o d e . • 0 to 10 MHz dc coupled. 100 Hz to 10 MHz ac coupled. 3. FREQ A Pulse Measurement Mode. • Channel A; 0 to 10 MHz 40 nsec minimum. 4. P E R A V G A M o d e . • 10 Hz (.1 sec) to 1 MHz (1 |j.sec). 5.
PARTS LIST
Model 5304A Replaceable Farts SECTION IX I) 5304A TIMER/COUNTER SUBSECTION VI REPLACEABLE P A R T S 9D-6-1. INTRODUCTION 9D-6-2. This subsection contains information for ordering replacement parts. Table 9D-6-1 lists parts used in the HP 5304A. Service Office (see lists in Section VI, 5300A Manual for addresses). Identify parts by their HewlettPackard part number. To obtain a part that is not listed, include: a. Instrument model number. 9D-6-3. ORDERING INFORMATION b. Instrument s ?rial number. 9D-6-4.
Model 5304 A Replaceable Parts Table 9D-6-1.
Model 5304A Replaceable Parts Table 9D-6-1. Replaceable Parts (Continued) Reference HP Part Number Designation A1R20 AM2I AM22 Al)<> » AM24 2100-1768 0757-028C 06 <>8-1444 •5757-0? 80 -#,•>8-3444 AIS/^, AM26 AU/7 AIR28 Al^7 06P 3 - I 00*> 068 3 - 1 0 0 5 0 6 8 3 - 680** 06P3-6805 068 3 - ? 7 1 5 A M JO A M 31 A |R 32 A H U A H 34 C6 8 3 - 2 71 5 *683-5105 C6P3-5I15 0683-51 I 5 P6P3-5105 AM35 A13 36 AM 37 A M 3t) A M 39 OG83-4715 069 3-1025 0683-4715 06*3-10?5 0683 2425 A|R
Model 5.304A Replaceable Parts Table 9D-6-1. Replaceable Parts (Continued) Reference HP Part Number Designation Qty Description Mfr Code Mfr Part Number A2J1 A2J2 A2SU A2R2 A21 3 1250-1163 1250-1163 06*13-91 * 5 C6P3-9145 06B3-9125 CONNf CTDP : R F 8NT. INPUT CONNf C T 0 R : R r RNC INPUT R : I X D COMP 9 1 OK OHM 5 * R : F X 0 COMP 9 1 0 K OHM 5 * R : f X 0 COMP 9 1 0 0 OHM 5 * 1/4H 1/4H 1/4H 2B4H0 28480 01121 01 121 01121 1250-1163 1250-1163 CH 0 | 4 5 f.
Model 5304A Manual Changes and Options S E C T I O N IX D 5304A T I M E R / C O U N T E R S U B S E C T I O N VII MANUAL CHANGES AND OPTIONS 9D-7-1. MANUAL CHANGES 9D-7-2. Section IX I) applies directly to Models 5304A Timer/Counter having serial prefix number 1212A00467. 9D-7-3. NEWER INSTRUMENTS 9D-7-4. As changes are made, newer instruments may have serial prefix numbers not listed in Section IX I).
Model 5304A Manual Changes and Options Table 9D-7-1. 5304A Section IX I) Timer/ Counter Title Page SECTION IX D TIMER/COUNTER 5304A S E R I A L P R E F I X : 1116A This section applies directly to HP Model 5304A, Timer/ Counters having serial prefix number 1116A, and must be inserted into the 5300A Measuring System Manual. NEWER INSTRUMENTS This Section with enclosed "Manual Changes Sheet(s)" applies directly to HP Model 5304A Timer/Counters having prefix numbers above 1118A.
Model r»:wMA Manual Changes and Options Table 9D-7-2. Table of Contents TABLE OF C O N T E N T S Section IX D Page 5304A Timer/Counter Subsection 9D-7-4 I GENERAL INFORMATION 91)1-1. Introduction 9D-1-2. Description 91)1-4. Purpose and Use of Section IX I) 91)1-6. Instrument Identification 91)1-8. Manual Changes and Options 91) 11 91)1-1 91)1-1 91)1-1 91)1-1 91)1-1 II INSTALLATION 9D-2-1. Unpacking and Inspection 9D-2-3. Storage and Shipment 9D-2-5. Installation and Removal of Plug-On 9D-2-8.
Model 5304A Manual Changes and Options Table 9]>-7-8.
Model 5304A Manual Changes and Options Table 90-7-4.
» El(6) — R50 EU7)— R56 El(8) — R58 E2( I) — R48 E2(6) — R55 E2(2) — R54 E2(7)— R5 I E2(3)—R52 E2(8)—R53 E2(4)—R59 E2(5) — R57 A1 » i A2 Bottom
NOTES 1 REFERENCE DESIGNATIONS WITHIN THIS ASSEMBLY AWE ABBREVIATED ADD ASSEMBLY NUMBER TO ABBREVIATION EOR COMPLETE DESCRIPTION 2 UNLESS OTHERWISE INDICATED RESISTANCE IN OHMS. CAPACITANCE IN PICOFARADS A2 ATTENUATOR BOARD ASSEMBLY ioiJO« sooo?. REFERENCE DESIGNATIONS NO PREFIX CI 15 CRI-IB El.? 01-19 R4 5 Rl - 70 SI.? [ LCVtL A I I i C V t l • 00E0l2]B 5*3 £ IAHIE OF ACTIVE COMPONENTS PART NUMBERS CRI 4 1901 0 3 7 6 CRVIZ.I6.I7 1901 - 0 0 4 0 CRlS.
Ai TiME N T E P V A L BOARD ASbEMBLY <0530« - 6000 • 016 0 0 0 0 0 0 m P/Q A I P l oxr 0 Oh - > 3 3 fir 012 Figure 9D-7-1.
1 SCHEMATC I DA IGRAMS
SECTION IX D 5304A TIMER/COUNTER S U B S E C T I O N VIII CIRCUIT DIAGRAMS 9D-8-1. GENERAL 9D-8-2. Subsection VIII contains: a. Schematic diagram notes are contained in Section V III of the 5300A portion of the manual. b. Instrument interconnection pin list of signals listing where they originate and where the go. c. Component locators, diagrams of assemblies.
Model 5304A Circuit Diagrams Table 9D-8-1. Instrument Interconnection List PIN NO. SIGNAL NAME 1 +5 V 2 -5 V 3 -17 V 4 GROUND DESCRIPTION Input DC from 5300A 5 * Fl 6 * 7 * "9" Goes low when the counter reaches 9% full scale. F2 The input signal to the time base gated by the control circuit. 8 9 * 10 * 11 INHIBIT Low signal forces the main gate flip-flop to the open position.
Model 5304A Circuit Diagrams Table 9D-8-1. Instrument Interconnection List PIN NO. SIGNAL NAME 35 NO CONNECTION p-36 DIGIT ADDRESS X ^37 DIGIT SELECT X [•*38 ^39 DIGIT ADDRESS Y p-40 DIGIT ADDRESS Z DIGIT SELECT Y *•41 DIGIT SELECT Z 42 NO CONNECTION 43 44 NO CONNECTION NO CONNECTION 44 45 NO CONNECTION NO CONNECTION DESCRIPTION Digit address code X, Y, Z from the display scanner indicates a digit being displayed.
Model 5304A Circuit Diagram Part of Figure 91) 8-1. 5304A Block Diagram PER AVERAGE FREO AUTO EXPONEN.CONTROL A INPUT EXP USA/B EXPONENT COUNTER TRANS U7A/B EXPONENT STORAGE T.I. A TO ANNUNC.. TB SELECT S2 FUNCTION SWITCH CHA AMP AUTO DECODERS D P I . 2 . 3.4.
EfflEI H* •^yywA.
NOTES 1 REFERENCE DESIGNATIONS WITHIN THIS ASSEMBLY ARE ABBREVIATED ADD ASSEMBt* NUMBER TO ABBREVIATION FOR COMPUTE DESCRIPTION 2 UNLESS OTHERWISE INDICATED RESISTANCE IN OHMS. CAPACITANCE IN PICOFARADS A2 A T T E N U A T O R BOARD ASSEMBLY 3»JO« - s o o o i ) R l f FRENCt DESIGNATIONS CI - i s CRI-18 El.
iSEMBLY 35S04-6000?> Ai TiME INTERVAL BOARD ASSEMBLY iOSJ04-60(X>o f >0 C55T | ItVtL i | | LtVtL » | I—&»' « • E E • 13 0 & 5*3 I 4ium S.CH* SCLCCTlO* LA y, OKJ'T tluix [OPf K C . 0 5 C t | <*"• tan P/O AlPt j ui»o ^ ?4 *4« * B J C U C t "C* 0 * >4* 2 ' * EM) • " [ M l 3>5 5i * '• run H 77^^-- ©o r—+>- r " 1 L> 40 ?*" A ?? re sfLtcT v l«4> 555 rr>M °** Figure 9D-8-1.
I MANUAL CHANCES
MANUAL CHANGES CHANGE DATE: December 15, 1981 This change supersedes all earlie* dated changes. *** Make all corrections listed under ERRATA before making other changes. MANUAL DESCRIPTION « « « * « « « « « « * * « * * « « » « * INSTRUMENT: 5304A * Timer-Counter Module * Operating and Service Manual * * * SERIAL PREFIX: *** Check following table for your instrument's serial prefix or series number and make listed change(s) to manual.
MANUAL CHANGES MODEL 5304A (05304-90006) Page 1 ERRATA Page 9D-2-1, 9D-3-2, 9D-3-6, and 9D-3-10: >Add the following: WARNING TO AVOID POSSIBILITY OF BODILY INJURY AND/OR EQUIPMENT DAMAGE, DO NOT CONNECT "HOT" SIDE OF POWER LINE TO INSTRUMENT CHASSIS GROUND. HEWLETT-PACKARD RECOMMENDS USING AN ISOLATION TRANSFORMER WHEN MEASURING AC LINE FREQUENCIES. ADDITIONALLY, DO NOT EXCEED THE INPUT VOLTAGE LIMITATIONS SPECIFIED IN TABLE 9D-1-1. Page 9D-3-8, Figure 9D-3-7.
MANUAL CHANGES MODEL 5304A (05304-90006) Page 2 CHANGE 1 (1704A) Page 9D-6-2, Table 9D-6-1, A1 (05304-60001) Replaceable Parts: >Add "SERIES 1704" to Description for A1 circuit board. >Change A1Q1 and A1Q2 from 05304-80001 to 1855-0213; DUAL JFET N-CHANNEL 2N5912; 28480; 1855-0213. >Change A1R19 and A1R20 from 2100-1768 (20-ohm) to 2100-2010; RESISTOR-VAR COMP 10-ohm 20%; 32997; 3329H-1-10R. . .• Page 9D-8-5, A1 (05304-60001) Schematic Diagram: >Add "SERIES 1704" at top of A1 (05304-60001) schematic.
MANUAL CHANGES MODEL 5304A (05304-90006) Page 3 CHANGE 5 (Serial Prefix No. 2024A) Page 9D-6-3 A2 (05304-60003), Replaceable Parts: >Change A2 (05304-60003) Series Number to 2024A. Page 9D-6-4 A2 (05304-60003)* Replaceable Parts: >Change A2S1, S2 and S3 from 3101-2383 to 3101-1313;SWITCH-SLIDE DPDT 0.5A 125V 28480; 3101-1313. CHANGE 6 Pages 90-6-3 and 9D-6-4, Table 9D-6-1. A2 (05304-60003) Replaceable Parts: >Change A2 SERIES number to 2148.
•* <1 MANUAL CHANGES MANUAL IDENTIFICATION ~ Model Number: 5 3 0 4 A Date Printed: A u g u s t 1 9 7 3 Part Number: 0 5 3 0 4 - 9 0 0 0 6 This supplement contains important information for correcting manual errors and for adapting the manual to instruments containing improvements made after the printing of the manual. To use this supplement: Make all ERRATA corrections. Make all appropriate serial number related changes indicated in the tables below.
Model No. 5304A 05304-90006 ERRATA (Cont'd): Page 9D-5-18, Table 9D-5-1, In Cabinet Performance Checks: Change Step e.4. to "COM'SEP-CHK to SEP". Change Step e.5. to "LEVEL (both channels) to position giving most stable display". Page 9D-6-2/3, Table 9D-6-1: * Change A1R19/20 to 2100-1985. Add MP 1 Part Number 1460-1312 Spring Clip. Add MP1 Part Number 0905-0479 Gasket. CHANGE 1 Page 9D-6-2, Table 9D-6-1: Add A1 Part Number 05304-20001 Board Blank RevD.
Model No. 5304A 05304-90006 CHANGE 3 (Cont'd): Page 9D-6-3, Table 9D-6-1: Change A1U15 to 1820-1112 INT CCT SN74LS74. Page 9D-8-5, A1 (05304-60001) Schematic Diagram: Add "NOTE 4" between A1Q1/A1Q3 and A1Q2/A1Q4. Add the following under NOTES: 4. "A1Q1/A1Q3 and A1Q2/A1Q4 ARE DUAL JFET N-CHAN UNITS WITH CENTER LEAD TO CASE CUT OFF". Change Ql-4 in TABLE OF ACTIVE COMPONENTS to Ql/q3, Q2/Q4 and HP PART NUMBER to 1855-0213. Page 9D-8-5: In table of active components: Change U15 to 1820-1112.