3. DIAGNOSTIC CHECK
Service Manual BSM-4100 3.33
3. Safety Check
There are the following 6 check items monitored by the NIBP safety circuit.
<Description and Acceptable Range of Check item>
Check item Description Range
PRESS 1 Pressure limiter in the adult mode 315 +/− 15 mmHg
PRESS 2 Pressure limiter in the neonatal mode 157.5 +/− 7.5 mmHg
TIME 1 Time limiter in the adult mode 176 - 180 sec ( >15 mmHg )
TIME 2 Time limiter in the neonatal mode 86 - 90 sec ( >5 mmHg )
TIME 3 Interval error in the adult mode >=10 sec ( >15 mmHg )
TIME 4 Interval error in the neonatal mode >=10 sec ( >5 mmHg )
NOTE
Since the TIME 1 to TIME 4 checks require a pressure generator which
supplies a pressure such as more than 5 mmHg or more than 15 mmHg
during a certain period such as 3 or 1.5 minutes or supplies the
pressure at a certain interval such as 10 seconds, do not perform these
checks locally.
Connect the manometer and hand bulb pump to the dummy cuff using the Y-shape
connector and connect the dummy cuff to the instrument with the air hose as
shown below.
For the PRESS 1, use the YN-901P air hose.
For the PRESS 2, use the YN-921P air hose.