3. DIAGNOSTIC CHECK
Service Manual BSM-4100 3.29
AUX I/F Check
This check is used to check the waveform signal output when the ZB-800PG/K
transmitter is connected to the instrument. The check program outputs a 15 Hz
sawtooth waveform to the transmitter. The radio telemetry receiver, such as a
cardiac telemetry system, receives the transmitted sawtooth waveform signal and
displays it on the screen.
Procedure to start the AUX I/F Check
1. To select the AUX I/F Check, touch the “AUX I/F CHECK” key at the DPU
CHECK column.
The instrument communicates with the transmitter and if it recognized the
connection with the transmitter, indicates “ZB CONNECTION” and outputs the
sawtooth waveform to the transmitter.
2. If there is no connection with the transmitter, the instrument indicates “ZB NO
CONNECTION”.
3. To return to the MANUAL CHECK MENU screen, touch the “RETURN” key.