Service manual
3. DIAGNOSTIC CHECK
3.28 Service Manual BSM-4100
AD TEST
This is not a check. The AD Test table shows the A to D converted values of all the
analog parameters such as ECG, SpO
2
and NIBP. The other parameters are shown
on the Multi Parameter Unit Check.
Each vital sign data are read from the A to D converter every 2 ms. The maximum
and minimum data of each vital sign data are displayed every 512 ms.
ECG1: ECG waveform data for heart rate STS1: 12 lead ECG status 1 data from the CPU on the ECG RESP3 board
ECG2: ECG waveform data for ZB-800P STS2: 12 lead ECG status 2 data from the CPU on the ECG RESP3 board
SpO2: Plethysmographic pulse wave data I: Lead I waveform data of 12 lead ECG
NIBP: NIBP measurement data II: Lead II waveform data of 12 lead ECG
RESP: Respiration waveform data V1: Lead V1 waveform data of 12 lead ECG
ECG3: ECG waveform data for display V2: Lead V2 waveform data of 12 lead ECG
MP1: Multi-parameter 1 socket input data V3: Lead V3 waveform data of 12 lead ECG
MP2: Multi-parameter 2 socket input data V4: Lead V4 waveform data of 12 lead ECG
MP3: Multi-parameter 3 socket input data V5: Lead V5 waveform data of 12 lead ECG
V6: Lead V6 waveform data of 12 lead ECG
Procedure to start the AD Test
1. To select the AD Test, touch the “AD TEST” key at the DPU CHECK column.
The program shows above table automatically.
2. To return to the MANUAL CHECK MENU screen, touch the “RETURN” key.