Service manual

1. GENERAL
Service Manual BSM-4100 1.17
Environment
Operating environment
Temperature: 10 to 40°C
Humidity: 30 to 90% RH (0 to 40°C, non-condensing)
Atmospheric pressure: 68 to 106 kPa
Storage environment
Temperature: 20 to +65°C
15 to +55°C (Recording paper)
Humidity: 15 to 95% RH (0 to 40°C, non-condensing)
Atmospheric pressure: 68 to 106 kPa
Dimensions and Weight (approximate)
Dimensions: 339 mm W × 375 mm H ×159 mm D
Weight: 9 kg (excluding battery)
Electromagnetic Compatibility
IEC60601-1-2: 1993 – Collateral Standard: Electromagnetic compatibility – Requirement and tests
Emissions: CISPR11 Group 1, Class B
Safety Standard
Safety standard: IEC 60601-1: 1988 Amendment 1: 1991, Amendment 2: 1995
IEC 6060-2-27: 1994 - Particular requirements for the safety of
electrocardiographic monitoring
IEC 60601-2-30: 1995 - Particular requirements for the safety of automatic cycling
in direct blood pressure monitoring equipment
IEC 60601-2-34: 1994 - Particular requirements for the safety of direct blood
pressure monitoring equipment
IEC 60601-2-49: 2001 - Particular requirements for the safety of multifunction
patient monitoring equipment
According to the type of protection against electrical shock:
CLASS I EQUIPMENT (AC Powered)
Internally Powered EQUIPMENT (BATTERY Powered)
According to the degree of protection against electrical shock
ECG, Respiration (impedance), IBP: Defibrillator-proof type CF applied part
Temperature, CO: CF applied part
NIBP: Defibrillator-proof type BF applied part
SpO
2
, Respiration (thermistor), CO
2
, FiO
2
: BF applied part
According to the degree of protection against harmful ingress of water:
IPX0 (ordinary EQUIPMENT)
According to the degree of safety of application in the presence of FLAMMABLE ANAESTHETIC MIXTURE WITH AIR,
OR WITH OXYGEN OR NITROUS OXIDE:
Equipment not suitable for use in the presence of FLAMMABLE ANAESTHETIC
MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE
According to the mode of operation: CONTINUOUS OPERATION