Service manual
1. GENERAL
Service Manual BSM-4100 1.15
CO
2
measuring accuracy:
TG-900P/TG-920P: ±3 mmHg (0 ≤ CO
2
≤ 10 mmHg)
±4 mmHg (10 < CO
2
≤ 40 mmHg)
±10% reading (40 < CO
2
≤ 100 mmHg)
(At 1 atmospheric pressure, air inspiration, no condensation)
TG-950P: ±2 mmHg (0 ≤ CO
2
≤ 40 mmHg)
±5% reading (40 < CO
2
≤ 70 mmHg)
±7% reading (70 < CO
2
≤ 100 mmHg)
(When no condensation)
Warm-up time: 5 s (minimum)
Response time
TG-900P: 160 ms (typical) for steps from 10 to 90%
TG-920P: 120 ms (typical) for steps from 10 to 90%
TG-950P: 60 ms (typical) for steps from 10 to 90%
Respiration rate counting range:
TG-900P/TG-920P: 3 to 150 breaths/min ±10%
TG-950P: 0 to 150 breaths/min ±10%
Respiration rate counting accuracy: ±2 breaths/min
CO
2
value display update cycle: Every 3 s or when alarm is generated
Alarm:
Upper limit range: 2 to 99 mmHg in 1 mmHg in 1 steps, OFF
Lower limit range: OFF, 1 to 98 mmHg in 1 steps
Apnea time: OFF, 5 to 40 s
Trendgraph
Trend parameters: Heart rate (or pulse rate), respiration rate, VPC rate, ST level, EVENT, apnea (time),
apnea (frequency), SpO
2
, NIBP (systolic, diastolic and mean), IBP (systolic,
diastolic and mean), temperature, CO
2
and FiO
2
Trend times: 1, 2, 4, 8, and 24 h
Vital Signs List
Parameters: Heart rate (or pulse rate), VPC rate, ST level, NIBP (systolic, diastolic and mean),
SpO
2
, IBP (systolic, diastolic and mean), respiration rate, temperature, CO
2
and
FiO
2
Number of files in list: Periodic vital signs list: 120
Entries in vital signs list at NIBP measurement: 120
List interval: Periodic vital signs list: 1, 5, 15, 30 or 60 minutes
Vital signs list at NIBP measurement: at NIBP measurement
Hemodynamics List (BSM-4103/4104/4113/4114 monitors only)
Number of files: 8
Full Disclosure (QM-421P flash memory card required)
Storage time: 48 hours
Alarm History
Number of files: 200