Service manual
1. GENERAL
1.12 Service Manual BSM-4100
Waveform display:
Display sensitivity: 10 mm/Ω ±25% (at ×1 sensitivity)
Sensitivity control: ×1/4, ×1/2, ×1, ×2, ×4
Respiration rate display update cycle: Every 3 s or when alarm is generated
Alarm:
Upper limit range: 2 to 150 breaths/min in 2 breaths/min steps, OFF
Lower limit range: OFF, 0 to 148 breaths/min in 2 steps
Apnea time: OFF, 5 to 40 s in 5 s steps
SpO
2
on BSM-4101/4103/4111/4113 Monitors
Measuring range: 50 to 100%
Pulse rate counting range: 0, 12 to 300 beats/min
SpO
2
accuracy: ±2 digits (80% ≤ SpO
2
≤ 100%)
±3 digits (50% ≤ SpO
2
< 80%)
SpO
2
display:
Pulse rate display update cycle: Every 3 s or when alarm is generated
Sync tone modulation: Change in 20 steps at 81 to 100% SpO
2
Waveform sensitivity: ×1/8, ×1/4, ×1/2, ×1, ×2, ×4, ×8 or AUTO
Alarm:
Upper limit range: 51 to 100% SpO
2
in 1% SpO
2
steps, OFF
Lower limit range: OFF, 50 to 99% SpO
2
in 1% SpO
2
steps
SpO
2
on BSM-4102/4104/4112/4114 Monitors
Measuring range: 1 to 100%
Pulse rate counting range: 0, 12 to 300 beats/min
SpO
2
accuracy: DS-100A: ±3% SpO
2
OXIBAND A/N: ±3% SpO
2
OXIBAND P/I: ±3% SpO
2
RS-10: ±3.5% SpO
2
D-YS: ±3% SpO
2
D-YSE: ±3.5% SpO
2
D-YSPD: ±3.5% SpO
2
D-25/D-25L: ±2% SpO
2
N-25: ±2% SpO
2
I-20: ±2% SpO
2
D-20: ±2% SpO
2
R-15: ±3.5% SpO
2
OXICLIQ I: ±2.5% SpO
2
OXICLIQ P: ±2.5% SpO
2
OXICLIQ A: ±2.5% SpO
2
OXICLIQ N: ±2.5% SpO
2
OXI1-2-3 A/N: ±2.5% SpO
2
OXI1-2-3 P/I: ±2.5% SpO
2
SpO
2
display:
Pulse rate display update cycle: Every 3 s or when alarm is generated
Sync tone modulation: Change in 20 steps at 81 to 100% SpO
2
Waveform sensitivity: ×1/8, ×1/4, ×1/2, ×1, ×2, ×4, ×8 or AUTO
Alarm:
Upper limit range: 51 to 100% SpO
2
in 1% SpO
2
steps, OFF
Lower limit range: OFF, 50 to 99% SpO
2
in 1% SpO
2
steps