Service manual

6. MAINTENANCE
6.12 Service Manual BSM-4100
Power
Power cord (and ground lead) are not damaged. Yes No
Instrument is firmly grounded to a dedicated grounding terminal. Yes No
Fuse is the correct rating. Yes No
Safety
Protective earth impedance is less than prescribed limit. Yes No
Earth leakage current is less than prescribed limits. Yes No
Enclosure leakage current is less than prescribed limits. Yes No
Patient leakage current is less than prescribed limits. Yes No
Withstand voltages of instrument are prescribed voltage and time limits or more. Yes No
Others
Line voltage is within the range of nominal voltage ±10%. Yes No
Instrument is correctly connected to other equipment. Yes No
All screws are tight. Yes No