Service manual
6. MAINTENANCE
6.10 Service Manual BSM-4100
Others
Item Check Procedure Action
Line voltage Check that the line voltage is within the range of
nominal voltage
±
10%.
Use only the line voltage within the
correct range.
Connection to other
equipment
Check that the instrument is correctly connected
to other equipment according to the operator’s
manual.
If there is a wrong connection, connect
the equipment to the instrument correctly
according to the operator’s manual.
Vibration or
resonant sound
Check that the instrument has no abnormal
vibration or resonant sound.
Remove the cause of abnormal vibration
or resonant sound.
Screws Check that there are no loose screws. If any screw is loose, tighten it.
Option check
Check that there is no error on the screen when
performing the optional card checks such as ATA
Card Check and Network Card Check on the
Manual Check Menu screen.
Remove the cause if an error occurs.
Installation
condition check
Check that the instrument is installed in a suitable
location according to the operator’s manual.
If the conditions are not suitable, improve
them.