Service manual

5. DISASSEMBLY AND ASSEMBLY
Service Manual BSM-4100 5.11
NOTE in Assembling the Instrument
To assemble the motherboard and all the analog boards, align the
projection bottom edge of each analog board with the bottom edge of
the inverted T shape holes on the motherboard (as shown below) so
that both side connectors on each analog board and the motherboard
can correctly catch each other. Then, depress the area of the
connectors, i.e. CNJ601, CNJ602, CNJ401, CNJ301 to 305.
Tilt the motherboard holder so that the bottom edge of the
motherboard holder can be inserted into the groove of the lower front
enclosure. Keep the insertion and lay the motherboard holder on the
motherboard.
3. Disconnect the air tube from the NIBP Measure board and remove the screw
which secures the NIBP Measure board to the DPU board. Remove the NIBP
Measure board from the DPU board.
Projection