Service manual
5. DISASSEMBLY AND ASSEMBLY
Service Manual BSM-4100 5.9
CAUTION in Assembling the Instrument
When you fasten the NIBP Safety board and analog output cable at the
ferrite core with the longer shaft screw, do not leave the analog output
cable on the three ICs (U802, U096 and U097) of the DPU board.
<Wrong> <Correct>
If the analog output cable is long and it has been left on the three ICs,
bind the cable with a cable binder as shown below so that the cable is
not left on the three ICs of the DPU board.
The longer shaft screw secures the NIBP
Safety board and ferrite core which holds
the analog output cable.
The unnecessary part of the analog output
cable is bound by a cable binder.